03 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Corrected System Performance
The specifications in this section apply to measurements made with the Keysight E5080A
network analyzer under the following conditions:
– No averaging applied to data
– Environmental temperature of 23 °C (± 3 °C) with less than 1 °C deviation from the
calibration temperature
– Response and isolation calibration performed
System dynamic range1
Specification (10 Hz IFBW)
Typical (3 Hz IFBW)
9 k to 100 kHz
100 k to 50 MHz
50 M to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
120 dB
129 dB
135 dB
132 dB
120 dB
137 dB
146 dB
152 dB
145 dB
135 dB
System dynamic range (specification and actual measurement data example)
1. The test port dynamic range is calculated as the difference between the test port rms noise floor and
the source maximum output power. The effective dynamic range must take measurement uncertainty
and interfering signals into account.