09 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Test port output power1
Description
Specification
Typical
Nominal power (preset power)
0 dBm
Range
9 k to 100 kHz
100 k to 50 MHz
50 M to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
—90 to 10 dBm
—90 to 12 dBm
—90 to 15 dBm
—90 to 12 dBm
—90 to 8 dBm
—110 to 12 dBm
—110 to 14 dBm
—110 to 17 dBm
—110 to 14 dBm
—110 to 12 dBm
Max leveled power
9 k to 100 kHz
100 k to 50 MHz
50 M to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
10 dBm
12 dBm
15 dBm
12 dBm
8 dBm
12 dBm
14 dBm
17 dBm
14 dBm
12 dBm
Resolution
0.01 dB
Level accuracy2
Sweep type: stepped
9 k to 50 kHz
± 2.0 dB
± 1.5 dB
50 k to 9 GHz
Sweep type: auto
9 k to 9 GHz
± 2.0 dB
Level linearity
—20 dBm to maximum power
Sweep type: stepped
Sweep type: auto
± 0.75 dB
± 1.0 dB
± 1.0 dB
—110 dBm to —20 dBm
Power sweep range
9 k to 100 kHz
100 k to 50 MHz
50 M to 6GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
—90 to 10 dBm
—90 to 12 dBm
—90 to 15 dBm
—90 to 12 dBm
—90 to 8 dBm
Test port output signal purity
Description
Specification
Typical
Harmonics (2nd or 3rd)
(at 0 dBm)
9 k to 2 GHz
2 G to 9 GHz
< –25 dBc
< –20 dBc
Non-harmonic spurious
(at 0 dBm)
< -30 dBc
1. Source output performance on all ports.
2. Power calibration using an external power meter improves level accuracy of the test port output power.
3. Level linearity given is relative to 0 dBm.