欢迎访问ic37.com |
会员登录 免费注册
发布采购

E4980A-002 参数 Datasheet PDF下载

E4980A-002图片预览
型号: E4980A-002
PDF下载: 下载PDF文件 查看货源
内容描述: [Precision LCR Meter]
分类和应用:
文件页数/大小: 38 页 / 1698 K
品牌: KEYSIGHT [ Keysight Technologies ]
 浏览型号E4980A-002的Datasheet PDF文件第1页浏览型号E4980A-002的Datasheet PDF文件第2页浏览型号E4980A-002的Datasheet PDF文件第4页浏览型号E4980A-002的Datasheet PDF文件第5页浏览型号E4980A-002的Datasheet PDF文件第6页浏览型号E4980A-002的Datasheet PDF文件第7页浏览型号E4980A-002的Datasheet PDF文件第8页浏览型号E4980A-002的Datasheet PDF文件第9页  
03 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Measurement functions  
Basic Specifications  
Measurement parameters  
– Cp-D, Cp-Q, Cp-G, Cp-Rp  
– Cs-D, Cs-Q, Cs-Rs  
– Lp-D, Lp-Q, Lp-G, Lp-Rp, Lp-Rdc  
– Ls-D, Ls-Q, Ls-Rs, Ls-Rdc  
– R-X  
– Z-θd, Z-θr  
– G-B  
– Y-θd, Y-θr  
– Vdc-Idc1  
Definitions  
Cp  
Cs  
Lp  
Ls  
D
Capacitance value measured with parallel-equivalent circuit model  
Capacitance value measured with series-equivalent circuit model  
Inductance value measured with parallel-equivalent circuit model  
Inductance value measured with series-equivalent circuit model  
Dissipation factor  
Q
Quality factor (inverse of D)  
G
Rp  
Rs  
Equivalent parallel conductance measured with parallel-equivalent circuit model  
Equivalent parallel resistance measured with parallel-equivalent circuit model  
Equivalent series resistance measured with series-equivalent circuit model  
Rdc Direct-current resistance  
R
Resistance  
X
Reactance  
Z
Impedance  
Y
Admittance  
qd  
qr  
B
Phase angle of impedance/admittance (degree)  
Phase angle of impedance/admittance (radian)  
Susceptance  
Vdc Direct-current voltage  
Idc Direct-current electricity  
Deviation measurement function: Deviation from reference value and percentage of  
deviation from reference value can be output as the result.  
Equivalent circuits for measurement: Parallel, Series  
Impedance range selection: Auto (auto range mode), manual (hold range mode)  
Trigger mode: Internal trigger (INT), manual trigger (MAN), external trigger (EXT),  
GPIB trigger (BUS)  
1. E4980A-001 is required.