09 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Relative accuracy  
Relative accuracy includes stability, temperature coefficient, linearity, repeatability, and  
calibration interpolation error. Relative accuracy is specified when all of the following  
conditions are satisfied:  
– Warm-up time: 30 minutes  
– Test cable length: 0 m, 1 m, 2 m, or 4 m (Keysight Technologies, Inc. 16048A/D/E)  
– A “Signal Source Overload” warning does not appear.  
When the test signal current exceeds a value in table 11 below, a “Signal Source  
Overload” warning appears.  
Table 11.  
Test signal voltage Test frequency  
Condition1  
≤ 2 Vrms  
> 2 Vrms  
≤ 1 MHz  
The smaller value of either 110 mA or  
130 mA - 0.0015 × Vac × (Fm / 1 MHz) × (L_cable + 0.5)  
70 mA - 0.0015 × Vac × (Fm / 1 MHz) × (L_cable + 0.5)  
> 1 MHz  
Vac [V]  
Fm [Hz]  
L_cable [m]  
Test signal voltage  
Test frequency  
Cable length  
– OPEN and SHORT corrections have been performed.  
– Bias current isolation: Off  
– The DC bias current does not exceed a set value within each range of the DC bias  
current  
– The optimum impedance range is selected by matching the impedance of DUT to the  
effective measuring range.  
– Under an AC magnetic field, the following equation is applied to the measurement  
accuracy.  
A x (1 + B x ( 2 + 0.5 / Vs))  
Where  
A:  
Absolute accuracy  
B:  
Vs:  
Magnetic flux density [Gauss]  
Test signal voltage level [Volts]  
|Z|, |Y|, L, C, R, X, G, and B accuracy (L, C, X, and B accuracies apply when Dx ≤ 0.1,  
R and G accuracies apply Qx ≤ 0.1)  
Relative accuracy Ae is given as:  
Equation 8.  
Ae = [Ab + Zs /|Zm| × 100 + Yo × |Zm| × 100 ] × Kt  
Impedance of DUT  
Basic accuracy  
Zm  
Ab  
Zs  
Yo  
Kt  
Short offset  
Open offset  
Temperature coefficient  
D accuracy  
D accuracy De is given as  
– when Dx ≤ 0.1  
Equation 9.  
De = ±Ae/100  
Measured D value  
Relative accuracies of |Z|, |Y|, L, C, R, X, G, and B  
Dx  
Ae  
1. When the calculation result is a negative  
value, 0 A is applied.  
– when Dx > 0.1, multiply De by (1 + Dx)