22 | Keysight | Infiniium 9000 Series Oscilloscopes - Data Sheet
Infiniium 9000 Series Performance Characteristics (Continued)
Acquisition
9104, 9254, 9404
9064
Maximum real-time sample rate
Memory depth per channel
Standard
Option 50M
Option 100
4 ch x 10 GS/s or 2 ch x 20 GS/s
4 ch x 5 GSa/s or 2 ch x 10 GSa/s
20 Mpts on 4 channels, 40 Mpts on 2 channels
50 Mpts on 4 channels, 100 Mpts on 2 channels
100 Mpts on 4 channels, 200 Mpts on 2 channels
200 Mpts on 4 channels, 400 Mpts on 2 channels
Option 200
Option 500
500 Mpts/250 Mpts on 4 channels, 1 Gpts/500 Mpts on 2 channels (single/repetitive mode)
Sampling modes
Real-time
Real-time with peak detect
Real-time with high resolution (user selectable to 9-, 10-, 11-, or 12-bits of resolution)
Real-time with roll mode (200 ms to 20 sec.)
Equivalent-time (1.0 ps fine interpolator resolution yields a maximum effective sample rate of 1,000 GSa/s)
Segmented memory (1 ps time stamp resolution between segments)
Maximum time between triggers is 562,950 seconds (6.5 days)
Re-arm time (minimum time between trigger events) is 4.5 μs with analog channels, 5.8 μs with digital channels on
Memory depth:
Max # of segments:
Filters
20 M
4096
50 M
8192
100 M
16384
200 M
32768
500 M
65536
1 G
131072
2 G
131072
Acquisition: Digital channels
Maximum real time sample rate
Maximum memory depth per channel
Minimum width glitch detection
Trigger: Scope channels
Trigger sources
2 GSa/s
128/64 Mpts with 2 GSa/s. 64/32 Mpts with sampling < 2 GSa/s (single/repetitive mode)
2 ns
Channel 1, channel 2, channel 3, channel 4, aux, and line
Sensitivity
1 MΩ input, edge trigger
50 Ω
DC to 500 MHz: 0.6 div
DC to 2 GHz, 0.5 div
2 GHz to 4 GHz: 1.0 div
DC to 700 MHz: 300 mVp-p
Auxiliary
Trigger level range
Channel 1, 2, 3 or 4
Auxiliary
Sweep modes
Display jitter (displayed trigger jitter) 1, 2
± 4 div from center screen (50 Ω)
± 8 div from center screen with max of ± 8 V (1 MΩ)
± 5 V (50 Ω up to 500 MHz with at least 500 mV signal swing)
Auto, triggered, single
9064 .50
9104 .35
9254 .50
9404 .40
Trigger holdoff range
Trigger actions
100 ns to 10 s fixed and random
Specify an action to occur, and the frequency of the action, when a trigger condition occurs
Actions include: e-mail on trigger and execute “multipurpose” user settings
Trigger coupling
1 MΩ: DC, AC, (10 Hz) low frequency reject (50 kHZ high pass filter), high frequency reject (50 kHz low pass
filter)
1� Internal edge trigger mode� Trigger threshold = fixed voltage at 50% level� The slew rate independent value in the formula represents the traditional trigger
jitter�
2� Display jitter example� At 100 mV/div typical noise values are 3�2 mV RMS for 9404 models, 2�5 mV RMS for 9254A models, and 2�4 mV RMS for 9104A
models� For slew rate of 500 mVpp sin wave with frequency equal to max analog bandwidth of each model, typical display jitter is �95 ps RMS for 9404A
models, �97 ps for 9254A models, and 1�7 ps RMS for 9104A models�