15 | Keysight | CX3300 Series Device Current Waveform Analyzer - Data Sheet
Current Sensor measurement range
Table 3. CX1101A Current Sensor, Single Channel, ± 40 V, 100 MHz, 40 nA - 1 A
Channel
RMS Noise 1
Maximum
Bandwidth (-3 dB)
Input Resistance Maximum Common
(Typical)
Mode Voltage
Range
10 A
10 mA
2 mA
3 MHz 2
15 mΩ
1 A
200 mA
20 mA
2 mA
0.2 mA
20 µA
3 µA
100 MHz
410 mΩ
50 Ω
±40 V
200 µA
20 µA
400 nA
40 nA
25 kHz
Table 4. CX1102A Current Sensor, Dual Channel, ± 12 V, 100 MHz, 40 nA - 1 A
Primary channel
Secondary channel
Maximum
Bandwidth (-3 dB)
Input Resistance Maximum Common
RMS Noise 1
Range
RMS Noise 1
(Typical)
Mode Voltage
Range
1 A
2 mA
0.2 mA
20 uA
2 uA
20 mA
2 mA
20 uA
100 MHz
410 mΩ
200 mA
20 mA
2 mA
3 uA
200 uA
500 nA
200 nA
400 nA 3
40 nA 3
±12 V
500 kHz
50 Ω
20 uA
20 mA 3
2 mA 3
8 uA 3
1 uA 3
200 uA 3
20 uA 3
90 kHz 3
25 kHz 3
50 Ω 3
Table 5. CX1103A Current Sensor, Low Side, 200 MHz, 100 pA - 20 mA
Channel
RMS Noise 1
Maximum Bandwidth
(-3 dB)
Maximum Offset
Current
Input Resistance Maximum Common
(Typical)
Mode Voltage
Range
20 mA
2 mA
5 uA
200 MHz
75 MHz
9 MHz
50 Ω
(50 Ω
±1 V
(50 Ω
±20 mA
±200 uA
±20 uA
1.5 uA
150 nA
25 nA
1.5 nA
150 pA
200 uA
20 uA
2 uA
Input ON)
4 Ω
Input ON)
±0.5 V
2.5 MHz
250 kHz
100 kHz
(50 Ω
(50 Ω
input OFF)
Input OFF)
200 nA
1. Noise measurement bandwidth = 20 MHz
2. At –4 dB.
3. Built-in low pass filter mode is ON.