14 | Keysight | CX3300 Series Device Current Waveform Analyzer - Data Sheet
The CX3300 series offers you the complete measurement
environment for visualizing wideband and low-level, previously
unmeasurable or undetectable, current waveforms.
The latest battery powered or energy harvested devices must further reduce power/
current consumptions. For repeatable and quantitative power reductions, it is necessary
to be able to visualize previously unmeasurable or undetectable wideband low-level
current waveforms. By adding the CX3300 series to our instrument lineup, Keysight
has provided a complete solution for low power device measurements that covers both
static/dynamic and voltage/current.
Figure 33. The CX3300 series can be an additional powerful instrument on your benchtop.
Table 1. CX3300 series mainframe key characteristics
Analog
bandwidth
options
Input
channel
Max. sample
rate
Memory depth
options
User installed
upgrade
Models
Dynamic range (ADC bits)
Digital channel
50 MHz, 100
MHz, 200 MHz
14 (high speed mode) and
16 (high resolution mode)
16 Mpts, 64
Mpts, 256 Mpts
Bandwidth,
Memory depth
CX3322A
CX3324A
2
4
1 GSa/s
1 GSa/s
No
50 MHz, 100
MHz, 200 MHz
14 (high speed mode) and
16 (high resolution mode)
16 Mpts, 64
Mpts, 256 Mpts
Bandwidth,
Memory depth
Yes
Table 2. Current sensors key characteristics
Models
Description
RMS noise @
20 MHz NBW
Max. measurable
current
DC measurement accuracy
(with mainframe) 1
Max. bandwidth
(standalone)
CX1101A
CX1102A 3
CX1103A
Current Sensor, Single Channel
Current Sensor, Dual Channel
Current Sensor, Low Side
40 nA
40 nA
150 pA
10 A 2
±(0.6 % + 0.3 %)
±(0.6 % + 0.3 %)
±(0.6 % + 0.3 %)
100 MHz
100 MHz
200 MHz
1 A
20 mA
1. 24 hrs after executing the user calibration. At 20 mA range.
2. With CX1206A sensor head
3. CX1102A occupies two input channels of mainframe