28 | Keysight | CX3300 Series Device Current Waveform Analyzer - Data Sheet  
CX1151A Passive Probe Interface Adapter characteristics  
Warranted specifications are denoted by **, and all others are supplemental characteristics. Specifications are valid after a 30-minute warm-up and 23 ± 5 ºC. All these  
characteristics are defined by 14-bit acquisition resolution unless otherwise stated.  
About the measurement accuracy  
Measurement accuracy can be affected by RF electro-magnetic field having the strengths greater than 3 V/m in the frequency range of 80 MHz to 2 GHz or 1 V/m in the  
frequency range of 2 GHz to 2.7 GHz. The extent of this effect depends upon how the instrument is positioned and shielded.  
Table 31. CX1151A characteristics overview  
Range  
8 V  
Noise (rms) 1  
5.0 mV  
Maximum bandwidth (-3 dB) 2  
DC offset range and resolution  
4 V  
2.8 mV  
±16 V, 16-bit resolution  
1.6 V  
0.4 V  
0.2 V  
0.08 V  
1.8 mV  
Voltage  
measurement  
300 MHz  
250 μV  
140 μV  
90 μV  
±0.8 V, 16-bit resolution  
1. Full bandwidth measured with mainframe (Option B20; 200 MHz bandwidth).  
2. Standalone bandwidth with an N2843A passive probe. The effective measurement bandwidth when connected to mainframe can be estimated by the  
following equation.  
1
BWeffective  
=
1
1
1
BWmainframe  
BWprobe  
2
2
BWadapter  
2
+
+
Table 32. CX1151A DC measurement accuracy 1  
Range  
With mainframe  
Standalone  
23 ±5 ºC  
TUSERCAL ±3 ºC, 24 hrs  
(High speed mode) 2  
TUSERCAL ±3 ºC, 24 hrs (High  
resolution mode) 3  
23 ±5 ºC  
8 V  
± (0.6% + 0.8%) **  
± (0.6% + 0.8%) **  
± (0.6% + 0.8%) **  
± (0.6% + 0.8%) **  
± (0.6% + 0.8%) **  
± (0.9% + 1.2%) **  
± (1.3% + 1.7%) **  
± (1.3% + 1.7%) **  
± (1.3% + 1.7%) **  
± (1.3% + 1.7%) **  
± (1.3% + 1.7%) **  
± (1.6% + 2.1%) **  
± (0.4% + 0.6%)  
± (0.8% + 0.6%)  
± (0.8% + 0.6%)  
± (0.4% + 0.6%)  
± (0.8% + 0.6%)  
± (0.8% + 0.6%)  
± (0.3% + 0.4%)  
± (0.5% + 0.4%)  
± (0.5% + 0.4%)  
± (0.3% + 0.4%)  
± (0.5% + 0.4%)  
± (0.5% + 0.4%)  
4 V  
1.6 V  
0.4 V  
0.2 V  
0.08 V  
Gain [% of readings] +  
Offset [% of range]  
Range with 10:1 passive probe 4  
TUSERCAL ±3 ºC, 24 hrs  
(High speed mode) 2  
TUSERCAL ±3 ºC, 24 hrs (High  
resolution mode) 3  
80 V  
40 V  
± (2.1% + 0.6%)  
± (1.5% + 0.6%)  
± (0.7% + 0.6%)  
± (1.7% + 0.6%)  
± (1.2% + 0.6%)  
± (0.4% + 0.6%)  
± (1.1% + 0.4%)  
± (0.8% + 0.4%)  
± (0.4% + 0.4%)  
± (0.9% + 0.4%)  
± (0.7% + 0.4%)  
± (0.3% + 0.4%)  
16 V  
4 V  
Gain [% of readings] +  
Offset [% of range]  
2 V  
0.8 V  
1. DC measurement condition: 20 ms averaged.  
2. After executing the User Calibration with mainframe. High speed mode (14-bit). Supplemental characteristics.  
3. After executing the User Calibration with mainframe. High resolution mode (16-bit). Supplemental characteristics.  
4. Passive probe used: N2843A.  