09 | Keysight | 89601B/BN-200 Basic Vector Signal Analysis, 89601B/BN-300 Hardware Connectivity, 89600 VSA Software - Technical Overview  
Software Features  
Basic VSA (Option 200)  
Note: The following features are independent of hardware platform used, unless otherwise noted.  
Time and waveform  
Time record characteristics  
In the 89600 VSA software, measurements are based on time records. A time record is a block  
of samples of the signal waveform from which time, frequency, and modulation domain data is  
derived.  
Data mode  
Zoom  
Two signal processing modes, baseband and zoom, affect the appearance and the duration of input  
waveforms displayed.  
Measurements are made with non-zero start frequency. Time domain display shows a complex envelope  
representation of the input signal, i.e. the magnitude and phase of the signal relative to the analyzer’s  
center frequency.  
Baseband  
Measurements begin at 0 Hz. The input signal is directly digitized and the waveform display shows the  
entire signal (carrier plus modulation), much as an oscilloscope would.  
Time record length (main time)  
Time sample resolution  
(Number of frequency points – 1)  
Span with RBW mode set to arbitrary, auto-coupled  
1/(k x span)  
Where:  
k = 2.56 for time data mode set to baseband  
k = 1.28 for all other modes (default) including zoom  
Span = Currently selected frequency span  
Time recording characteristics  
In recording (time capture) mode the incoming waveform is captured gap-free into high-speed time  
capture memory. This data may then be replayed at full or reduced speed, saved to mass storage, or  
transferred to another software application.  
When time analyzing the captured waveform, users may adjust measurement  
span and center frequency in order to zoom in on a signal, as long as the new measurement span lies  
entirely within the originally captured span.  
Time recording memory size  
Resolution bandwidth (RBW)  
RBW values  
Memory size is dependent on the hardware used. See hardware specifications for more information.  
The range of available RBW choices is a function of the selected frequency span and the number of cal-  
culated frequency points. Users may step through the available range in a 1-3-10 sequence or directly  
enter an arbitrarily chosen bandwidth.  
Range  
< 1 Hz to > 0.287 x Max Span  
RBW shape factor  
The window choices below allow the user to optimize the RBW shape as needed for best amplitude ac-  
curacy, best dynamic range, or best response to transient signal characteristics.  
Selectivity  
0.41  
Passband flatness  
0.01 dB  
Rejection  
> 95 dBc  
> 125 dBc  
> 31 dBc  
> 13 dBc  
> 92 dBc  
> 89 dBc  
> 73 dBc  
Flat top  
Gaussian top  
Hanning  
0.25  
0.68 dB  
0.11  
1.5 dB  
Uniform  
0.0014  
0.26  
4.0 dB  
Blackman-Harris  
Kaiser-Bessel  
Gaussian  
0.83 dB  
0.26  
0.83 dB  
0.22  
0.83 dB