21 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications (Continued)
Dual/quad electrical
Dual/Quad electrical channel modules
Channel options
N1045A 1
04F/04M
N1055A 1
52F/52M
02F/02M
32F/32M
34F/34M
54F/54M
(number of channels; F = female; M = male)
Number of channels 2
2
4
2
4
2
4
Electrical channel bandwidth
Transition time (characteristic)
(10 to 90% calculated from TR = 0.35/BW)
Channel-to-channel skew range
RMS noise
20/35/45/60 GHz
17.5/10/7.8/5.8 ps
35 GHz4, 5
10 ps
35 GHz 4, 5
35/50 GHz 5
10 /7 ps
35/50 GHz 5
± 100 ps
± 150 ps
Characteristic
Maximum
275/425/500/750 µV
950 µV (60 GHz)
600 µV
730 µV
600/750 µV
950 µV
Scale factor (per division)
Minimum
1 mV/division
1 mV/division
Maximum
100 mV/division
100 mV/division
DC accuracy (single marker)
± 2 mV ± 4% of (reading-channel ± 2 mV ± 4% of (reading-channel offset), (50 GHz)
offset), (60 GHz)
DC offset range (referenced from center of
screen)
± 500 mV
± 500 mV
Input dynamic range (relative to channel offset) ± 400 mV
± 400 mV
+2 V/–1 V
50 Ω
Maximum input signal
Nominal impedance
± 2 V (+16 dBm)
50 Ω
Reflections (for 30 ps rise time)
Electrical input 3
20%
20%
1.85 mm (female or male option) 2.92 mm (female or male option) 1.85 mm (female or male option)
TDR system
TDR system
Oscilloscope/TDR performance
Normalized characteristics
(Mainframe with 54754A module)
Rise time
40 ps nominal
Adjustable from larger of 17 ps or 0.08 x time/div
< 25 ps normalized
≤ ± 1% after 1 ns from edge
≤ + 5%, –3% < 1 ns from edge
0.00 V ± 2 mV
Maximum: 5 x time/div
≤ 0.1%
TDR step flatness
Low level
High level
± 200 mV +2 mV
TDR system
N1055A - 3xx
N1055A - 5xx
(Mainframe with N1055A module)
(35 GHz bandwidth)
(50 GHz bandwidth)
Without TDR calibration
< 18 ps
With TDR calibration
Adjustable from 15 ps,
characteristic
Without TDR calibration
< 7 ps
With TDR calibration
Adjustable from 6 ps,
characteristic
Step rise/fall time (transmitted 6)
1. Module is supported by 86100D DCA-X mainframe and later.
2. Upgradable from two channel to four channel after purchase (return to Keysight).
3. Connector style is the same on all channels and is selected at time of order.
4. Upgradable from 35 to 50 GHz after purchase (return to Keysight).
5. Tuned to be –3 dB (± measurement uncertainty) at stated bandwidth(s) using NIST traceable swept-sine test system.
6. Defined as the transition time at the output of the remote head. It is calculated by deconvolving the receiver transition time from the measured transition
time when the remote head is terminated with a short. Measured on a negative TDR step.