06 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Specifications (Continued)
Mainframe specifications (Continued)
Option STR (standard trigger)
Option ETR (enhanced trigger)
Trigger modes
Internal trigger 1
External direct trigger 2
Limited bandwidth 3
Full bandwidth
Free run
Free run
DC to 100 MHz
DC to 3.2 GHz
DC to 100 MHz
DC to 3.2 GHz
External divided trigger
PatternLock
Module bay trigger 7
Jitter
N/A
N/A
N/A
3 to 13 GHz (15 GHz, 32 GHz with Option PTB)
50 MHz to 13 GHz (50 MHz to 15 GHz)
Yes, supported
Characteristic
< 1.0 ps RMS + 5*10E-5 of delay setting 4
1.5 ps RMS + 5*10E-5 of delay setting 4
N/A
1.2 ps (750 fs Option PTB) RMS for time delays less than 100 ns 6
1.7 ps (1.2 ps Option PTB) RMS for time delays less than 100 ns 6
1.2 ps (750 fs) (50 MHz to 32 GHz) in PTB mode 8: 200 fs
(2.4 to 4 GHz) 120 fs (4 to 9 GHz) 90 fs (9 to 44 GHz)
Maximum
Option PTB
Trigger sensitivity
200 mVpp (sinusoidal input or 200 ps
minimum pulse width)
200 mVpp sinusoidal input: 50 MHz to 13 GHz, (to 32 GHz with
Option PTB)
Trigger configuration
Trigger level adjustment
Edge select
–1 to + 1 V
Positive or negative
Normal or High sensitivity
AC coupled
N/A
N/A
Hysteresis 5
Trigger gating
Gating input levels
(TTL compatible)
Gating delay
Disable: 0 to 0.6 V, Enable: 3.5 to 5 V
Pulse width > 500 ns, period > 1 μs
Disable: 27 μs + trigger period +
Maximum time displayed
Enable: 100 ns
Trigger impedance
Nominal impedance
Reflection
50 Ω
< 10% for 100 ps rise time
3.5 mm (male)
Connector type
Maximum trigger signal
2 V peak-to-peak
1. The freerun trigger mode internally generates an asynchronous trigger that allows viewing the sampled signal amplitude without an external trigger signal
but provides no timing information unless a precision timebase is used in Eye/Mask mode. Freerun is useful in troubleshooting external trigger problems.
2. The sampled input signal timing is recreated by using an externally supplied trigger signal that is synchronous with the sampled signal input.
3. The DC to 100 MHz mode is used to minimize the effect of high frequency signals or noise on a low frequency trigger signal.
4. Measured at 2.5 GHz with the triggering level adjusted for optimum trigger.
5. High Sensitivity Hysteresis mode improves the high frequency trigger sensitivity but is not recommended when using noisy, low frequency signals that may
result in false triggers without normal hysteresis enabled.
6. Slew rate ≥ 2 V/ns.
7. The Module Bay Trigger routes trigger signals from the module’s rear panel to the mainframe. 86100D-ETR is recommended when using a DCA module
equipped with a rear-panel trigger circuit. Examples include 54754A, 83496x, and 86108A/B modules. If operating these modules in an 86100D with Option
STR, an external cable (such as P/N 5062-6690) must be connected from the module’s front panel trigger/clock output to the 86100D’s trigger input.
8. In precision timebase (PTB) operation with 750 mV input (500 mV input, 9 to 44 GHz).