12 | Keysight | 16860A Series Portable Logic Analyzers - Data Sheet
16860A Series Interfaces
Peripheral interfaces
Display
One 15-pin XGA connector and one DisplayPort connector
Audio ports
Line in, line out, mic in
USB ports
Two 3.0 ports on rear, two 2.0 ports on front, two 2.0 ports on rear
Connectivity interfaces
LAN
Connector
10Base-T, 100Base-T, 1000Base-T
RJ-45
Interface with external instrumentation
Trigger or arm external devices or receive signals that can be used to arm measurement hardware within the logic analyzer with trigger in/out
Trigger in
Input
Rising edge or falling edge
Action taken
Input signal level
Threshold level
When received, the logic analyzer takes the actions described in the trigger sequence step
± 6 V max
Selectable: ECL, LVPECL, LVTTL, PECL, TTL
User defined (± 5 V in 50 mV increments)
Minimum signal amplitude
Connector
200 mV
BNC
Input resistance
Trigger out
3.2 kΩ nominal
Trigger
Select one event from the following as the trigger out event: logic analyzer trigger, flag 1, flag 2, flag 3 or
flag 4
Mode
Feedthrough
Polarity
Active high
Output signal
VOH (output high level) 2.0 V min
VOL (output low level) 0.5 V max
Signal load
50 Ω (For good signal quality, the trigger out signal should be terminated in 50 Ω to ground)
Connector
BNC
External reference clock in
10 MHz ± 0.01%
Signal swing
Connector
Minimum 200 mVp-p swing, Maximum 5 Vp-p swing
BNC