欢迎访问ic37.com |
会员登录 免费注册
发布采购

16861A-032 参数 Datasheet PDF下载

16861A-032图片预览
型号: 16861A-032
PDF下载: 下载PDF文件 查看货源
内容描述: [Portable Logic Analyzers]
分类和应用:
文件页数/大小: 22 页 / 2683 K
品牌: KEYSIGHT [ Keysight Technologies ]
 浏览型号16861A-032的Datasheet PDF文件第3页浏览型号16861A-032的Datasheet PDF文件第4页浏览型号16861A-032的Datasheet PDF文件第5页浏览型号16861A-032的Datasheet PDF文件第6页浏览型号16861A-032的Datasheet PDF文件第8页浏览型号16861A-032的Datasheet PDF文件第9页浏览型号16861A-032的Datasheet PDF文件第10页浏览型号16861A-032的Datasheet PDF文件第11页  
07 | Keysight | 16860A Series Portable Logic Analyzers - Data Sheet  
16860A Series Logic Analyzer Specifications and Characteristics  
State (synchronous) sampling mode  
16861A  
16862A  
16863A  
16864A  
Channels  
34  
68  
102  
(96 data and  
6 clock)  
136  
(128 data and  
8 clock)  
(32 data and  
2 clock)  
(64 data and  
4 clock)  
Sampling option: Single clock  
Clock (clock is on Pod 1)  
Clock qualifiers  
1
1
0
1
3
0
1
4
0
1
4
1
Reset qualifier  
Sampling option: Multiple clocks  
Clocks or clock qualifiers  
Reset qualifier  
2
0
4
0
4
0
4
0
Clock channels can be used as data channels.  
The state sampling clock mode specifies how the clock inputs are used for sampling. The  
availability of these state sampling clock modes depends on the state sampling option that you  
select.  
Master - All pods sampled by the master clock definition.  
In single clock mode, only the clock signal on Pod 1 can be used.  
In multiple clocks, either a single clock signal can be used or a combination of clocks  
can be used.  
Dual sample - In the dual sample clock mode, you can capture two samples per clock  
edge with two different threshold offsets and separate sampling positions. These separate  
threshold offsets and sampling positions allow you to set independent thresholds and  
sampling positions for Read and Write in DDR/LPDDR captures and for Rising and Falling  
edge in general-purpose data captures.  
Master/slave - Master pod is sampled on master clock and slave pod is sampled on slave  
clock, but the captured data of both slave and master clocks is saved together when the  
master clock occurs.  
Demux - Data being probed by one pod is demultiplexed into the logic analyzer memory  
that is normally used for two pods. The demultiplex mode uses the master and slave clocks  
to demultiplex the data.  
350 MHz  
(Base configuration)  
700 MHz  
(Option 700)  
350 MHz  
(Base configuration)  
Sampling option  
Available clock modes  
Single clock  
Master  
Single clock  
Master  
Multiple clocks  
Master  
Dual sample  
Dual sample  
Master/slave  
Demux  
Maximum state data  
rate (spec) 1  
Captures data up to  
Captures data up to  
Captures data up to  
350 Mbps on either edge 700 Mbps on either edge 700 Mbps on any  
of a clock up to 350 MHz of the clock up to  
700 MHz  
combination of multiple  
clocks up to 350 MHz  
Captures data up to  
Captures data up to  
700 Mbps on both edges 1400 Mbps on both  
of a clock up to 350 MHz edges of the clock up to  
700 MHz  
Maximum state clock  
frequency  
Minimum state clock  
frequency  
350 MHz  
700 MHz  
350 MHz  
0 MHz  
12.5 MHz (single edge),  
6.25 MHz (both edges)  
12.5 MHz (single edge),  
6.25 MHz (both edges)  
Minimum time between 1430 ps  
active clock edges  
715 ps  
1430 ps  
700 MHz  
Maximum trigger  
sequencer speed  
700 MHz  
1400 MHz  
1. Specification (spec): Represents warranted performance of a calibrated instrument that has been stored for a  
minimum of 2 hours within the operating temperature range of 5 to 40 °C, unless otherwise stated, and after  
a 45-minute warm-up period. The specifications include measurement uncertainty.