Table 4. Corrected system performance with type-N device connectors,
85092C electronic calibration (ECal) module
Network analyzer:
Calibration module:
E5071C
85092C (Type-N, 50 Ω) Electronic calibration (ECal)
module
full 2-port
Calibration:
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C
( 5 °C ) with ꢀ 1 °C deviation from calibration temperature, isolation calibration performed
Specification (dB)
300 kHz to
10 MHz
10 MHz to
3 GHz
3 GHz to
6 GHz
6 GHz to
8.5 GHz
Description
Directivity
45
52
52
47
Source match
Load match
36
44
41
36
41
47
44
39
Reflection tracking
Transmission tracking
0.100
0.055
0.040
0.039
0.060
0.068
0.070
0.136
Transmission uncertainty (specification)
Magnitude
10
Phase
100
10
1
300 K Hz to 10 MHz
300 kHz to 10 MHz
10 MHz to
3
G Hz
G Hz
G Hz to 8.5 G Hz
10 MHz to
3
G Hz
G Hz
G Hz to 8.5 G Hz
3
6
G Hz to 6
3
6
G Hz to 6
1
0.1
S11 = S22 = 0; Cal power = -10 dBm; Meas power = -10 dBm
S11 = S22 = 0; Cal power = -10 dBm; Meas power = -10 dBm
0.1
0.01
10
0
-10 -20 -30 -40 -50 -60 -70 -80 -90
Transmission coefficient (dB)
10
0
-
10
-
20
-
30
-
40
-
50
-
60
-
70
-
80
-90
Transmission coefficient (dB)
Reflection uncertainty (specification)
Magnitude
Phase
0.05
10
8
300 kHz to 10 MHz
10 MHz to 3 G Hz
3 G Hz to 6 G Hz
300 kHz to 10 MHz
10 MHz to 3 G Hz
3 G Hz to 6 G Hz
0.04
0.03
0.02
0.01
0
6 G Hz to 8.5 G Hz
6 G Hz to 8.5 G Hz
S21 = S12 = 0; Cal power = -10 dBm; Meas power = -10 dBm
S21 = S12 = 0; Cal power = -10 dBm; Meas power = -10 dBm
6
4
2
0
0
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
Reflection coefficient (linear)
8