Corrected system performance with calibration kit
Option 230/235/240/245/260/265/280/285/430/435/440/445/460/465/480/485
Table 3. Corrected system performance with type-N device connectors,
85032F calibration kit
Network analyzer:
Calibration kit:
Calibration:
E5071C
85032F (Type-N, 50 Ω)
full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C
( 5 °C ) with ꢀ 1 °C deviation from calibration temperature, isolation calibration performed
Specification (dB)
9 kHz to
300 kHz to 10 MHz to 3 GHz to 6 GHz to
Description
300 kHz
10 MHz
49
3 GHz
46
6 GHz
40
8.5 GHz
38
Directivity
49
Source match
Load match
41
41
40
36
35
49
49
46
40
37
Reflection tracking
Transmission tracking
0.011
0.027
0.011
0.015
0.021
0.018
0.032
0.056
0.054
0.088
Transmission uncertainty (specification)
Magnitude
10
Phase
100
10
1
9 KHz to 300 KHz
300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
9 KHz to 300 KHz
300 KHz to 10 MHz
10 MHz to 3 GHz
6 GHz to 8.5 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
1
0.1
S11
= S22 = 0; Cal power = -10 dBm ; Meas power = -10 dBm
S11
= S22 = 0; Cal power = -10 dBm ; Meas power = -10 dBm
0.01
10
0.1
0
–10 –20 –30 –40 –50 –60 –70 –80 –90
Transmission coefficient (dB)
10
0
-10 -20 -30 -40 -50 -60 -70 -80 -90
Transmission coefficient (dB)
Reflection uncertainty (specification)
Magnitude
Phase
10
8
0.05
9 KHz to 300 KHz, 300 KHz to 10 MHz
10 MHz to 3 GHz
9 KHz to 300 KHz, 300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
3 GHz to 6 GHz
0.04
6 GHz to 8.5 GHz
6 GHz to 8.5 GHz
6
0.03
4
0.02
0.01
2
S21 = S12 = 0; Cal power = -10 dBm; Meas power = -10 dBm
S21 = S12 = 0; Cal power = -10 dBm; Meas power = -10 dBm
0
0
0
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
Reflection coefficient (linear)
7