Measurement Throughput Summary
Cycle time for measurement completion1, 2 (ms)
Table 32. Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485
Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: OFF
Number of points
51
201
401
1601
Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth
Uncorrected
2-port cal
4-port cal
3.4
6.3
13
4.9
9.3
19
7.1
14
28
20
41
92
Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected
2-port cal
4-port cal
3.7
6.6
13
5.2
10
21
7.3
15
32
20
42
94
Start 1 GHz, stop 1.2 GHz, 1 kHz IF bandwidth
Uncorrected
2-port cal
4-port cal
54
202
401
800
397
1563
3125
6247
107
213
792
1581
Start 100 kHz, stop 3 GHz, 500 kHz IF bandwidth
Uncorrected
2-port cal
4-port cal
8.7
17
34
11
21
42
12
24
49
25
48
102
Start 100 kHz, stop 3 GHz, 100 kHz IF bandwidth
Uncorrected
2-port cal
4-port cal
8.8
17
34
11
22
42
13
24
49
26
54
118
Start 100 kHz, stop 3 GHz, 1 kHz IF bandwidth
Uncorrected
2-port cal
4-port cal
56
206
410
818
404
1583
3163
6324
112
222
805
1608
1. Supplemental performance data.
2. Measured with firmware version A.09.30.