Table 31. Analyzer environmental specifications and dimensions
Description
General characteristics
Operating environment
Temperature
+5 °C to +40 °C
Error-corrected temperature range
23 °C ( 5 °C) with < 1 °C deviation from calibration temperature
20% to 80% at wet bulb temperature < +29 °C (non-condensation)
0 to 2,000 m (0 to 6561 feet)
Humidity
Altitude
ꢀibration
0.21 G maximum, 5 Hz to 500 Hz
Non-Operating storage environment
Temperature
–10 °C to +60 °C
Humidity
20% to 90% at wet bulb temperature < +40 °C (non-condensation)
Altitude
0 to 4572 m (0 to 15,000 feet)
0.5 G maximum, 5 Hz to 500 Hz
See next page.
ꢀibration
Dimensions
Weight (net)
Option 230/240/260/280 (2-port)
Option 235/245/265/285 (2-port)
Option 430/480/440/460 (4-port)
Option 435/445/465/485 (4-port)
Option 2D5/2K5 (2-port)
18.2 kg
18.3 kg
19.9 kg
20.0 kg
19.8 kg
21.8 kg
Option 4D5/4K5 (4-port)