Performance Characteristics (continued)  
Power Requirements  
Line Voltage Range  
54621A/D, 54622A/D, 54624A: 100 - 240 VAC ± 10ꢂ, CAT II, automatic selection  
54641A/D, 54642A/D: 100-240 VAC, 50/60 Hz, CAT II, automatic selection; 100-132 VAC, 440 Hz, CAT II,  
automatic selection  
Line Frequency  
Power Usage  
54621A/D, 54622A/D, 54624A: 47 to 440 Hz  
54641A/D, 54642A/D: 50/60 Hz, 100-240 VAC; 440 Hz, 100-132 VAC  
54621A/D, 54622A/D, 54624A: 100 W max  
54641A/D, 54642A/D: 110 W max  
Environmental Characteristics  
Ambient Temperature  
Humidity  
Operating -10 ꢀC to +55 ꢀC; Non-operating -51 ꢀC to +71 ꢀC  
Operating 95ꢂ RH at 40 ꢀC for 24 hr; Non-operating 90ꢂ RH at 65 ꢀC for 24 hr  
Operating to 4,570 m (15,000 ft); Non-operating to 15,244 m (50,000 ft)  
Altitude  
Vibration  
HP/Agilent class B1 and MIL-PRF-28800F; Class 3 random  
1
Shock  
HP/Agilent class B1 and MIL-PRF-28800F; (operating 30 g, / sine, 11-ms duration, 3 shocks/axis along major axis.  
2
Total of 18 shocks)  
Pollution Degree2  
Indoor Use Only  
Normally only dry non-conductive pollution occurs.  
Occasionally a temporary conductivity caused by condensation must be expected.  
This instrument is rated for indoor use only  
Other  
Installation Categories  
CAT I:Mains isolated  
CAT II:Line voltage in appliance and to wall outlet  
Regulatory Information:  
Safety  
IEC 61010-1:1990+A1:1992+A2:1995 / EN 61010-1:1994+A2:1995  
UL 3111  
CSA-C22.2 No. 1010.1:1992  
Supplementary Information  
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive  
89/336/EEC, and carries the CE-marking accordingly. The product was tested in a typical configuration with  
HP/Agilent test systems.  
*
Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 ꢀC from firmware calibration temperature.  