Performance Characteristics (continued)
Display System
Display
7-inch raster monochrome CRT
Throughput of Analog Channels 25 million vectors/sec per channel with 32 levels of intensity
Resolution
255 vertical by 1000 horizontal points (waveform area) 32 levels of gray scale
Waveform intensity on front panel. Vectors on/off; infinite persistence on/off 8 x 10 grid with continuous intensity control
Key-specific help in 11 languages displayed by pressing and holding key or softkey of interest
Time and date (user setable)
Controls
Built-in Help System
Real Time Clock
Measurement Features
Automatic Measurements
Voltage (analog channels only)
Time
Measurements are continuously updated. Cursors track current measurement
Peak-to-Peak, Maximum, Minimum, Average, Amplitude, Top, Base, Overshoot, Preshoot, RMS (DC)
Frequency, Period, + Width, - Width and Duty Cycle on any channels
Rise time, Fall time, X at Max (Time at max volts), X at Min (Time at min volts), Delay, and Phase on analog channels only.
Counter
Built-in 5 digit frequency counter on any channel. Counts up to 125 MHz
Threshold Definition
Cursors
Variable by percent and absolute value; 10ꢂ, 50ꢂ, 90ꢂ default for time measurements
Manually or automatically placed readout of Horizontal (X, ∆X, 1/∆X) and Vertical (Y, ∆Y)
Additionally digital or analog channels can be displayed as binary or hex values
Waveform Math
One function of 1-2, 1*2, FFT, differentiate, integrate.
Source of FFT, differentiate, integrate: analog channels 1 or 2 (or 3 or 4 for 54624A), 1-2, 1+2, 1*2
FFT
Points
Fixed at 2048 points
Source of FFT
Window
Analog channels 1 or 2 (or 3 or 4 on 54624A only), 1+2, 1-2, 1*2
Rectangular, Flattop, Hanning
-70 to -100 dB depending on averaging
In dBV, dBm@50 Ω
Noise Floor
Amplitude Display
Frequency Resolution
Maximum Frequency
0.097656/(time per div.)
102.4/(time per div.)
*
Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 ꢀC from firmware calibration temperature.
18