HV-SPGU (high voltage semiconductor pulse generator unit) module specification  
Specifications  
Pulse/DC output voltage and accuracy  
Output voltage (Vout) 50 Ω load  
Open load  
Number of output channels:  
2 channels per module  
–20 V to ꢁ20 V  
–40 V to ꢁ40 V  
Modes: pulse, constant, and freerun  
Accuracy¹  
Open load  
(0.5 % ꢁ 50 mV)  
0.2 mV ( 10 V range)  
0.8 mV ( 40 V range)  
0.4 mV ( 10 V range)  
1.6 mV ( 40 V range)  
SMA  
Standard pulse mode:  
Two level pulse  
• Three level pulse per one channel  
• Pulse period: 30 ns to 10 s  
Amplitude resolution 50 Ω load  
Open load  
Output connectors  
Source impedance  
Short circuit current  
Delay range: 0 s to 9.99 s  
Delay resolution: 2.5 ns (minimum)  
Output count: 1 to 1,000,000  
Voltage monitor minimum sampling  
period: 5 μs  
50 Ω²  
800 mA peak (400 mA average³)  
(5 % ꢁ 20 mV)  
Overshoot/  
50 Ω load  
pre-shoot/ringing4  
Output limit  
1. At 1 µs after completing transition.  
2. Typical ( 1 %)  
Monitoring over current limit  
Trigger output:  
Level: TTL  
3. This value is specified under the following condition: [(Number of installed HV-SPGUs) ꢀ 0.2 A]  
ꢁ [DC current output by all modules (including HV-SPGUs)] < 3.0 A  
Timing: Synchronized with pulse period  
Trigger width:  
4. Following the specified condition with transition time.  
Pulse period ꢀ 1/2 (pulse period ≤ 10 µs)  
Maꢀimum 5 µs (pulse period > 10 µs)  
Pulse range and pulse parameter¹  
Frequency range  
Pulse period  
0.1 Hz to 33 MHz  
30 ns to 10 s  
10 ns  
SPGU supplemental information  
Programmable range  
Resolution  
Minimum  
Accuracy  
Programmable range  
Resolution  
Pulse width jitter: 0.001 % ꢁ150 ps  
Pulse period jitter: 0.001 % ꢁ150 ps  
Maꢀimum slew rate: 1000 V/µs (50 Ω load)  
Noise: 10 mVrms (at DC output)  
100 ns³  
1 % ( 0.01 % ²)  
10 ns to (period – 10 ns)  
2.5 ns (Tr and Tf ≤ 8 µs)  
10 ns (Tr or Tf > 8 µs)  
50 ns (25 ns typical)³  
(3 % ꢁ 2 ns)  
Width  
Advanced feature:  
Voltage monitor: The HV-SPGU has a  
voltage monitor function to measure the  
voltage at the DUT terminal.  
Minimum  
Accuracy  
Programmable range  
Transition time5  
(Tr and Tf)  
8 ns to 400 ms  
Measurement accuracy (open load):  
(0.1 % of reading ꢁ 25 mV)  
Measurement resolution: 50 μV  
Note: Specified at 1 PLC (20 ms = (5 μs  
sample ꢁ 5 μs interval) ꢀ 2000 samples.)  
Voltage compensation: The HV-SPGU  
can measure the impedance of DUT and  
adjust the output voltage according to  
the DUT impedance.  
Resolution  
2 ns (Tr and Tf ≤ 8 µs)  
8 ns (Tr or Tf > 8 µs)  
< 15 ns³  
20 ns (Vamp ≤ 10 V)  
30 ns (Vamp ≤ 20 V)  
60 ns (Vamp > 20 V)  
–5 % to 5 % ꢁ 10 ns (Vamp ≤ 10 V)  
–5 % to 5 % ꢁ 20 ns (Vamp ≤ 20 V)  
< 100 µs  
Minimum (typical)  
Minimum  
Accuracy  
Output relay  
Open/close  
switching time4  
1. Unless otherwise stated, all specifications assume a 50 Ω termination.  
ALWG (arbitrary linear  
2. Typical minimum. This is supplemental information.  
waveform generator) function  
3. This is specified at Vamp ≤ 10 V.  
4. The time it takes the open state relay to open or close.  
5. The time from 10 % to 90 % of Vamp which is the amplitude of output pulse.  
Arbitrary linear waveform generator  
(ALWG) mode:  
• Output compleꢀ waveform per one  
channel of HV-SPGU  
• Define multi-level pulse and multi-pulse  
waveform including open state pulse  
with ALWG GUI editor  
Pulse waveform  
Tperiod  
Twidth  
Tdelay  
Vpeak  
• Sequential pulse waveform from  
user-defined pulse waveform  
• 1024 points per one channel  
90%  
90%  
Vamp  
10%  
Tr  
10%  
Tf  
• Programmable timing range:  
10 ns to 10 s, 10 ns resolution  
Vbase  
11