Table 1. Maximum Ratings(1)
Rating
Symbol
P1max
Tstg
Value
2000
Unit
kPa
°C
Maximum Pressure(2) (P2 ≤ 1 Atmosphere)
Storage Temperature
–40 to +125
–40 to +125
Operating Temperature
TA
°C
1. Maximum Ratings apply to Case 867 only. Extended exposure at the specified limits may cause permanent damage or degradation to the
device.
2. This sensor is designed for applications where P1 is always greater than, or equal to P2. P2 maximum is 500 kPa.
Table 2. Operating Characteristics (VS = 5.0 Vdc, TA = 25°C unless otherwise noted, P1 > P2. Decoupling circuit shown in
Figure 4. required to meet electrical specifications.)
Characteristic
Symbol
POP
VS
Min
0
Typ
—
Max
500
5.25
10
Unit
kPa
Pressure Range(1)
Supply Voltage(2)
Supply Current
4.75
—
5.0
7.0
0.20
4.70
4.50
—
Vdc
IO
mAdc
Vdc
Zero Pressure Offset(3)
Full Scale Output(4)
Full Scale Span(5)
Accuracy(6)
(0 to 85°C)
(0 to 85°C)
(0 to 85°C)
(0 to 85°C)
Voff
VFSO
VFSS
—
0.088
4.587
—
0.313
4.813
—
Vdc
Vdc
—
2.5
%VFSS
mV/kPa
ms
Sensitivity
V/P
tR
—
9.0
1.0
0.1
20
—-
Response Time(7)
—
—-
Output Source Current at Full Scale Output
Warm-Up Time(8)
IO+
—
—-
mAdc
ms
—
—
—-
1. 1.0 kPa (kiloPascal) equals 0.145 psi.
2. Device is ratiometric within this specified excitation range.
3. Offset (Voff) is defined as the output voltage at the minimum rated pressure.
4. Full Scale Output (VFSO) is defined as the output voltage at the maximum or full rated pressure.
5. Full Scale Span (VFSS) is defined as the algebraic difference between the output voltage at full rated pressure and the output voltage at the
minimum rated pressure.
6. Accuracy (error budget) consists of the following:
• Linearity:
Output deviation from a straight line relationship with pressure over the specified pressure range.
• Temperature Hysteresis: Output deviation at any temperature within the operating temperature range, after the temperature is cycled to
and from the minimum or maximum operating temperature points, with zero differential pressure applied.
• Pressure Hysteresis:
Output deviation at any pressure within the specified range, when this pressure is cycled to and from the
minimum or maximum rated pressure, at 25°C.
• TcSpan:
• TcOffset:
Output deviation over the temperature range of 0° to 85°C, relative to 25°C.
Output deviation with minimum rated pressure applied, over the temperature range of 0° to 85°C, relative to 25°C.
• Variation from Nominal: The variation from nominal values, for Offset or Full Scale Span, as a percent of VFSS, at 25°C.
7. Response Time is defined as the time for the incremental change in the output to go from 10% to 90% of its final value when subjected to a
specified step change in pressure.
8. Warm-up Time is defined as the time required for the device to meet the specified output voltage after the pressure has been stabilized.
Table 3. Mechanical Characteristics
Characteristics
Weight, Basic Element (Case 867)
Typ
Unit
4.0
grams
MPX5500
Sensors
Freescale Semiconductor
2