1 push < 1s
1 push > 1s
Original State
(Auto/Manual)
1 push > 1s
1 push < 1s
〝REL〞active.
〝REL〞blink.
The relative value
The DN value
push HOLD
push HOLD
〝REL〞and〝H〞active.
〝REL〞and〝H〞blink.
1 push < 1s
Hold the relative value
The DN value
Note: It’s possible that the relative value exceeds 11,000 or –11,000 counts, but never
exceeds 19,999 or –19,999 counts. Such relative values are displayed. The LCD shows
OL in REL mode only if DN or DN+K is more than 10,999 counts or the relative value is
more than 19,999.
2.5 Max/Min + HOLD
The meter displays the maximum or minimum value of the input in Max/Min mode.
When Max/Min is pressed for the first time, the meter displays the maximum value.
When Max/Min is pressed again, the meter displays the minimum value. When
Max/Min is pressed for the third time, the meter displays current value. The meter
returns to normal operation if Max/Min is pressed and held for longer than one second.
Pressing HOLD in Max/Min mode makes the meter stop updating the maximum or the
minimum value.
1 push<1s
Original state
(Automatic/manual)
1 push>1 sec
1 push>1s
1 push>1 sec
1push<1s
1push<1s
“MIN” active
min value
“Max” “MIN” blink
current value
“MAX” active
max value
1push<1s
Push HOLD
Push HOLD
Push HOLD
“MAX”“MIN” “H” blink
hold current value.
“MIN” “H” active
hold min value.
“MAX” “H” active
hold max value.
1push<1s
1push<1s
1push<1s
v1.2
17
11/09/06