ES51964(6600counts)
Dual Display
6.3. Operating Timing
ES51964 incorporates a dual slope ADC with four phases: ZI, AZ, INT and DINT. The
timing of each phase is listed below. The Bar-graph ADC is 10X faster than slow ADC.
*Voltage / Ohm /ADP / Manual Current (for clampmeter) measurement:
Phase
ZI
Lo-speed
60ms
Hi-speed
6ms
AZ
INT
DINT
24ms
100ms
166ms
2.4ms
10ms
16.6ms
*Current mode for multimeter / Current mode for clampmeter
Phase
ZI
Lo-speed
60ms
Hi-speed
6ms
AZ
INT
DINT
24ms
1000ms
166ms
2.4ms
100ms
16.6ms
*Capacitance measurement:
Note: Cap. (Clamp mode) measurement time is double for 6.600mF-66.00mF ranges
Range
6.600nF
66.00nF
660.0nF
6.600μF
66.00μF
660.0μF
6.600mF
66.00mF
Total Measurement Time
0.35sec
0.35sec
0.35sec
0.35sec
0.70sec
1.40sec
1.40sec(max)
7.0sec(max)
*Frequency + Duty cycle mode measurement
In range 66.00Hz~66MHz, the measurement cycle = 1.05 sec
P.S. In the frequency measurement with auto mode, if the range is changed, the internal
clock rate will increase ten times and the new measurement cycle becomes 1/10 times of
the original cycle until the range is stable.