ES51925
3000 Counts Smart DMM
4. Miscellaneous
The conditions, which the meter turns on the buzzer, include:
(1) Changing measurement mode generates one beep.
(2) Pressing any of the push functions generates one beep, if the function is valid.
(3) Power on and re-power on generate one beep.
(4) Input overflow in voltage and current mode generates one beep every 0.3
seconds (or 3.33 beeps per second.)
(5) Continuity (diode) check generates a continuous 2KHz beep whenever the
measurement is less then 30Ω(30mV)
(6) Auto power off or power down by SEL key generates a 2KHz beeper that lasts
for 1.5 seconds.
The following figures show the output waveform from the BUZOUT pin.
0.5 mS
(a) Continuous 2KHz beep
0.15 sec
0.3 sec
(b) 3.33 beep/sec
Ver 2.9
16
10/06/25