CDB5530U  
are displayed in the Conversion Data box, as well as the Channel (always “1” for CS5530) and Overflow  
indicator bits for each conversion. The software will halt the data collection process when the user presses  
the Stop button.  
Configuration Register: In the Configuration Register box, the contents of the Configuration register are  
displayed, and can be modified by typing a hexadecimal value in the Hexadecimal: box, or by changing  
any of the values below the Hexadecimal: box to the desired settings.  
Note: When changing the value of the reset system bit to “1” (RS, bit 29 in the Configuration register), the part will  
be reset, and all registers will return to their default values. It is a good idea to select the Update Registers button  
after performing a reset to update the screen with the new register values.  
Offset / Gain Registers: In the Offset Registers / Gain Registers box, the contents of the offset and gain  
registers are displayed in hexadecimal. These registers can all be modified in the Calibration window,  
which is accessed when the user selects the Calibration Window button.  
Calibration Window: The Calibration window (Figure 6) permits the user to access the calibration regis-  
ters of the CS5530 and also perform system offset or gain calibrations. Prior to performing calibrations  
however, the Configuration register should be modified to configure the part for the desired operating  
mode (see Section 2.2.4, "Configuration Window" on page 13 for more details). Note that offset calibration  
should be performed before gain calibration to ensure accurate results. Refer to the CS5530 data sheet  
for more details.  
Figure 6. CDB5530U Calibration Window  
System Offset Calibrations:  
1) Attach an input signal to the AIN± inputs that you want to calibrate as zero. Note that the grounded  
input must be within the common mode range of the amplifier for proper operation.  
2) Activate the system offset calibration by pressing the System OFFSET button.  
3) When the calibration is complete, the screen will automatically update the offset register with the  
new value.  
System Gain Calibrations:  
1) Attach a full-scale input to the AIN± inputs of the CDB5530U.  
2) Activate the system gain calibration by pressing the System GAIN button.  
3) When the calibration is complete, the screen will automatically update the gain register with the  
new values.  