where +12V or +15V is supplied on traditional ADC774s.  
Pin 11, the –12V or –15V supply input on traditional  
ADC774s, is used only as a logic input on the ADS774.  
There is a resistor divider internally on pin 11 to reduce that  
input to a correct logic level within the ADS774, and this  
resistor will add 10mW to 15mW to the power consumption  
of the ADS774 when –15V is supplied to pin 11. To  
minimize power consumption in a system, pin 11 can be  
simply grounded (for Emulation Mode) or tied to +5V (for  
Control Mode.)  
If adjustment is required, connect the converter as shown in  
Figure 10. Sweep the input through the end-point transition  
voltage (0V + 1/2LSB; +1.22mV for the 10V range, +2.44mV  
for the 20V range) that causes the output code to be DB0 ON  
(HIGH). Adjust potentiometer R1 until DB0 is alternately  
toggling ON and OFF with all other bits OFF. Then adjust  
full scale by applying an input voltage of nominal full-scale  
minus 3/2LSB, the value which should cause all bits to be  
ON. This value is +9.9963V for the 10V range and +19.9927V  
for the 20V range. Adjust potentiometer R2 until bits DB1-  
DB11 are ON and DB0 is toggling ON and OFF.  
There are no other modifications required for the ADS774 to  
function with a single +5V supply.  
CALIBRATION PROCEDURE—BIPOLAR RANGES  
If external adjustments of full-scale and bipolar offset are  
not required, replace the potentiometers in Figure 11 by  
50, 1% metal film resistors.  
CALIBRATION  
OPTIONAL EXTERNAL FULL-SCALE  
AND OFFSET ADJUSTMENTS  
If adjustments are required, connect the converter as shown  
in Figure 11. The calibration procedure is similar to that  
described above for unipolar operation, except that the offset  
adjustment is performed with an input voltage which is  
1/2LSB above the minus full-scale value (–4.9988V for the  
±5V range, –9.9976V for the ±10V range). Adjust R1 for  
DB0 to toggle ON and OFF with all other bits OFF. To  
adjust full-scale, apply a DC input signal which is 3/2LSB  
below the nominal plus full-scale value (+4.9963V for ±5V  
range, +9.9927V for ±10V range) and adjust R2 for DB0 to  
toggle ON and OFF with all other bits ON.  
Offset and full-scale errors may be trimmed to zero using  
external offset and full-scale trim potentiometers connected  
to the ADS774 as shown in Figures 10 and 11 for unipolar  
and bipolar operation.  
CALIBRATION PROCEDURE—  
UNIPOLAR RANGES  
If external adjustments of full-scale and offset are not  
required, replace R2 in Figure 10 with a 501% metal film  
resistor and connect pin 12 to pin 9, omitting the other  
adjustment components.  
®
13  
ADS774  