FIGURE 5. Optional Offset Voltage Trim.  
Step 2: Output stage trim (RA = RB = 20M, RC = 100k,  
RF = 0, RX = 100k, RK = , R1 and R2 connected);  
A1 low, A2 low gain. Adjust R2 for 0V ±1mV or  
desired setting at VOUT, pin 15 (±110mV approxi-  
mate total range).  
pass section (0.05Hz cutoff) is formed by the 1µF capacitor  
and 3Mresistor which are connected in series between the  
output demodulator and the inverting input of the output  
stage amplifier. The low-pass section (100Hz cutoff) is  
formed by the 68Mresistor and 22pF capacitor located in  
the feedback loop of the output stage. The diodes provide for  
quick recovery of the high-pass filter to overvoltages at the  
input. The 100kpot and the 100Mresistor allow the  
output voltage to be trimmed to compensate for increased  
offset voltage caused by unbalanced impedances seen by the  
inputs of the output stage amplifier.  
NOTE: Other circuit component values can be used with  
valid results.  
APPLICATIONS  
ECG AMPLIFIER  
In many modern electrocardiographic systems, the patient is  
not grounded. Instead, the right-leg electrode is connected to  
the output of an auxiliary operational amplifier as shown in  
Figure 7. In this circuit, the common-mode voltage on the  
body is sensed by the two averaging resistors, R1 and R2,  
inverted, amplified, and fed back to the right-leg through  
resistor R4. This negative feedback drives the common-  
mode voltage to a low value. The body’s displacement  
current id does not flow to ground, but rather to the output  
circuit of A3. This reduces the pickup as far as the ECG  
amplifier is concerned and effectively grounds the patient.  
Although the features of the circuit shown in Figure 6 are  
important in patient monitoring applications, they may also  
be useful in other applications. The input circuitry uses an  
external, low quiescent current op amp (OPA177 type)  
powered by the isolated power of the input stage to form a  
high impedance instrumentation amplifier input (true three-  
wire input). R3 and R4 give the input stage amplifier of the  
3656 a noninverting gain of 10 and an inverting gain of –9.  
R1 and R2 give the external amplifier a noninverting gain of  
1 + 1/9. The inputs are applied to the noninverting inputs of  
the two amplifiers and the composite input stage amplifier  
has a gain of 10.  
The value of R4 should be as large as practical to isolate the  
patient from ground. The resistors R3 and R4 may be se-  
lected by these equations:  
The 330k, 1W, carbon resistors and diodes D1 - D4 provide  
protection for the input amplifiers from defibrillation pulses.  
R3 = (R1/2) (VO/VCM) and R4 = (VCM – VO)/id  
The output stage in Figure 6 is configured to provide a  
bandpass filter with a gain of 22.7 (68M/3M). The high-  
(–10VVO +10V and –10V VCM +10V)  
®
11  
3656  