欢迎访问ic37.com |
会员登录 免费注册
发布采购

AXRB9000_18 参数 Datasheet PDF下载

AXRB9000_18图片预览
型号: AXRB9000_18
PDF下载: 下载PDF文件 查看货源
内容描述: [High Stability Low Noise Rubidium Frequency Reference in 19” rack (2 HU) with up to 3 outputs]
分类和应用:
文件页数/大小: 5 页 / 237 K
品牌: AXTAL [ ADVANCED XTAL PRODUCTS ]
 浏览型号AXRB9000_18的Datasheet PDF文件第1页浏览型号AXRB9000_18的Datasheet PDF文件第3页浏览型号AXRB9000_18的Datasheet PDF文件第4页浏览型号AXRB9000_18的Datasheet PDF文件第5页  
Designed and  
Manufactured  
in Germany  
Parameter  
min.  
typ.  
max. Unit Condition  
Nominal output frequency  
Frequency stability  
10.000  
MS  
±0.5  
MHz  
STD  
±0.5  
±0.05  
±0.5  
±1.3  
HS  
Option 1 (max.)  
ppb steady state  
±0.03 ppb after 3 months operation  
vs. operating temperature range  
Long term (aging) per month  
Long term (aging) 1st year  
Long term (aging) over 10 years  
Frequency adjustment range  
Mechanical Frequency Control  
RF output  
±0.1  
±0.05  
±0.5  
±1.3  
±0.2  
±1  
ppb  
ppb  
after 3 months operation  
after 3 months operation  
±1.5  
ppb  
By trimmer access (Note 3)  
RF output ports  
1 Rubidium direct output  
3 splitted outputs  
Sine wave  
See block diagram and  
enclosure drawing  
Signal waveform  
Load RL  
50  
±5%  
Rubidium direct output  
Splitted outputs  
Output level per port  
+12  
+7  
+14  
+9  
dBm  
dBm  
dB  
Isolation between splitted ports  
Harmonics  
Spurious  
30  
40  
-40  
-80  
dBc  
dBc  
Phase noise  
Short-term stability (ADEV)  
See table 1  
MS  
3·10-11  
1·10-11  
3·10-12  
Option 2  
STD  
HS  
Option 1 (max.)  
@ τ = 1 sec  
@ τ = 10 sec  
8·10-11  
3·10-11  
8·10-12  
1·10-11  
3·10-12  
1·10-12  
@ τ = 100 sec  
< ±0.5 ppb  
Warm-up time @ +25°C (Note 2)  
Retrace  
15  
min  
±0.05 ppb  
V
1 h after 24 hrs OFF  
COMM interface  
IEC 60320-1 / C14  
Monitor logic voltage  
AC Supply voltage VS  
5
230  
100  
50  
240  
60  
V
AC Supply input frequency  
Power consumption  
Hz  
W
°C  
50  
+50  
Operating temperature range  
0
Enclosure (see drawing) (WxDxH)  
Drawing number  
483x250x88  
AXZ10.01097.02  
BNC female  
mm  
Color “black”  
RF Connectors  
@ Rear plate  
@ Rear plate  
Communication Connector  
9-Pin D-Sub male  
with jack posts  
Weight  
Notes:  
8
kg  
1. Terminology and test conditions are according to IEC60679-1 and MIL-PRF-55310, unless otherwise stated  
2. Warm-up indicator at front plate. Indicator ON when Rubidium locked.  
3. Trimmer accessible at front plate. Adjustment range sufficient for 10 years operation.  
Absolute Maximum Ratings  
Parameter  
min.  
max.  
Unit  
Condition  
AC Supply Voltage VS  
90  
47  
260  
63  
V
AC Supply input frequency  
AC Supply input current  
Splitter input level (SPLIT IN)  
Storage Temperature  
Hz  
A
2
Fuse accessible at rear plate  
+20  
+70  
dBm  
°C  
-20  
GmbH & Co. KG  
AXRB9000 Rev.3  
Drawing D0  
Page 2(5)  
www.axtal.com  
fon: +49 (6261) 939834  
fax: +49 (6261) 939836  
E-mail: info@axtal.com  
Roemerring 9  
D-74821 Mosbach  
Germany  
 复制成功!