9
PC card slot for type-III cards
A PC card-type GP-IB or
RS-232C interface is used for
this slot as well.
Waveform monitor (color TFT LCD)
External control
terminal
DC power inlet
Connect the optional DC
POWER ADAPTER 8439
AC power inlet
Floppy-disk drive
Logic probes input terminals
Analog input unit mount
Thermal printer: 110 mm in width, 8 dots/mm
8835-01 (use 2ch input type modules) 8835-01 (use 4ch ANALOG UNIT 8946)
Basic specifications
Memory functions
Plug-in input modules
Max. 4 analog chʼs + 16 logic chʼs Max. 8 analog chʼs + 16 logic chʼs
(Isolated analog channels, isolated input and
frame, logic has common GND)
Plug-in input modules
100μs to 5min/DIV, 20 ranges or external sampling, time axis
resolution 100 points/DIV, time axis zoom: ×2 to ×10 in 3 stages,
compression: 1/2 to 1/2,000 in 10 stages
Input type/number
of channels
Time axis
(Isolated analog channels, isolated input and
frame, logic has common GND)
1/100 of time axis ranges (minimum sampling period 1μs)
Max. 500kS/s (minimum sampling period 2μs)
Sampling rate
MEM (high-speed recording)
REC (real-time recording)
RMS (50/60Hz, or DC only)
REC & MEM (Additional functions provided by the optional FUNCTION UP DISK 9540-01)
FFT (Additional functions provided by the optional FUNCTION UP DISK 9540-01)
External sampling
Measurement
functions
Settable in 1 DIV steps, 20 to 40,000 DIV*1
Recording length
Pre-trigger
1
*
Depending on the number of channels in use.
Can record data from before the trigger point, 0 to 100% or -95% of
recording length; 15 settings
1 MS/second (1μs, all channels simultaneously)
External sampling (500kS/second, 2μs)
Maximum
sampling rate
waveform processing*2, waveform parameter processing, waveform
averaging*2, memory segmentation (up to 255 segments)*2, logging
(numerical printout), X-Y waveform plot, voltage axis zoom ×2 to ×
10, 3 settings, compression 1/2, zoom, variable display, graph
4 Mwords total
(12 analog bits + 4 logic bits) × 4 Mwords/channel (1 channel used) to (12 analog
Memory capacity
Other functions
bits + 4 logic bits) × 500 kwords/channel (8 channels used)
superimposition, waveform judgment function*2
PC Card Type III slot × 1: up to 1GB (Flash ATA)
2
*
Additional functions provided by the optional FUNCTION UP DISK 9540-01
Data storage media 3.5" Floppy disk drive × 1: 1.44MB, 1.2MB, 720KB, MS-DOS format
File format: Binary, text, BMP
Recorder functions (time axis waveform and X-Y format)
10ms to 1 hour/DIV with 17 ranges, time axis resolution 100 points/
DIV, time axis compression: 1/2 to 1/50 in 5 stages
Clock and setting conditions: battery life of at least 10 years
Waveform data: battery life of at least 1 hour after system power is
turned OFF (at 2 minutes after power-on)
Backup functions
(at 25°C)
At 10ms to 200ms/DIV, printing in real time is not possible, but waveform data are stored in
Time axis
memory and can be monitored on screen. Data are stored for 2,000 divisions before the end
of measurement. At recording length settings other than "Continuous", the printer can be
used simultaneously, for follow-up printing of waveforms.
Mini-jack 3.5 mm in dia.: Trigger input/output
Terminal board: External start, stop, print input/sampling input,
decision output
External control
connectors
1μs to 100ms; 6 settings (selectable from 1/100 or less of time axis)
Sampling rate
Settable in 1 DIV steps, 20 to 2,000 DIV*3, or "Continuous"*4
GP-IB, RS-232C, LAN
Interfaces (option)
Note: Use one of the following: RS-232C CARD 9557, GP-IB CARD 9558, LAN CARD
At X-Y format: only continuous for X-Y plotting
Recording length
(HIOKI-tested)
3
4
*
Measure all channels.
When time 10 ms − 200 ms/DIV and printer is ON, continuous is not available.
*
Operation: +5°C (41°F) to +40°C (104°F), 35% to 80% rh
Storage: -10°C (14°F) to +50°C (122°F), 20% to 90% rh
Environmental conditions
(no condensation)
100μs; fixed (dot), 100μs to 25ms (line)
X-Y sampling period
X-Y axis resolution
40dots/DIV (display), 80dots (horizontal) × 80 dots (vertical)/DIV (printer)
Compliance
standard
Safety: EN61010
EMC: EN61326, EN61000-3-2, EN61000-3-3
Store data for most recent 2,000 DIV in memory. Backward
scrolling and re-printing available.
Waveform memory
100 to 120V AC or 200 to 240V AC (50/60 Hz)
10 to 28V DC (use the DC POWER ADAPTER 9439 : option)
Power
requirements
logging (numerical printout), additional recording (recording is resumed
without overwriting previous data), voltage axis magnification ×2 to ×10;
3 settings, compression 1/2; 1 setting, variable display.
120VA, max. for 100V AC (approx. 60VA with the printer off)
70VA, max. for 12V DC (approx. 30VA with the printer off)
Other functions
Power consumption
(when using two units of 8936)
(approx. 110VA with the printer off)
170VA, max. for 100V AC
RMS Recorder Function (for 50/60 Hz and DC)
Power consumption
(when using two units of 8940)
(approx. 50VA with the printer off)
Note: When using the F/V UNIT 8940 with 12V DC power, the printer can only be used for
80VA, max. for 12V DC
5s to 1 hr/DIV; 9 settings, time axis compression 1/2 to 50; 5 settings
Time axis
up to 2 channels.
200μs fixed (20 rms datas/s)
Sampling rate
RMS calculation accuracy
Dimensions and
mass
285mm (11.22in)
(158.73oz.) (main unit only)
W × 220mm (8.66in) H × 132mm (5.20in) D, 4.5kg
±3% f.s.
Settable in 1 DIV steps, 20 to 2,000 DIV*5, or "Continuous"
Recording length
Waveform memory
Instruction Manual × 1, Power cord × 1, Printer paper 1, Protective
×
5
*
Measure all channels.
Supplied
accessories
cover 1, Roll paper attachment 2, PC card protector 1, Application
×
×
×
Store data for most recent 2,000 DIV in memory. Backward
scrolling and re-printing available.
Disk (Wave Viewer Wv, Communication Commands table)
1
×
Print/display section
logging (numerical printout), additional recording (recording is resumed
without overwriting previous data), voltage axis magnification ×2 to ×10;
3 settings, compression 1/2; 1 setting, variable display.
Other functions
6.4 inch TFT color LCD, with English/Japanese selector (640 × 480 dots)
Display
110 mm (4.33 in) × 30 m (98.4 ft), thermal paper roll
Recording paper
Recording width
Auxiliary Functions
10 divisions for full scale, 1 DIV = 10 mm (0.39 in) (80 dots)
Printing of settings including input range, trigger time, etc, cursor
10 rows/mm (250 rows/in) * 20 rows/mm (500 rows/in with the memory
recorder's smooth print function
Paper feed density
Recording speed
measurement, scaling, free comment input, screen hard copy,
registration of setting conditions (eight conditions), start condition
retention, auto setup, auto saving, remote control, auto ranging,
view function, online help, key lock, list printing, etc.
General
Max. 25 mm/s (0.98 in/s)
Trigger functions
Scaling: Translation of amplitude gradation only
Variable: Arbitrary setting of the upper and lower limit of the
waveform display range
CH1 to CH8 (analog), CHA to CHD (logic), external, timer, manual
(either ON or OFF for each source), logical AND/OR of sources
Trigger sources
Scaling
Level: Digital setting of voltage. Triggered when set value is
exceeded in UP or DOWN direction.
Allows precision adjustment of input voltage.
Vernier function
Average value, effective (rms) value, peak to peak value, maximum
value, time to maximum value, minimum value, time to minimum
value, period, frequency, rise time, fall time, area value, X-Y area
value, and standard deviation.
Window: When entering or exiting a level range defined by upper
or lower limit
Voltage drop: Only for AC power lines. Triggered when the peak
voltage falls below setting value
Waveform parameter
calculation
(in MEM function)
Trigger types
(analog)
RMS level: Only for DC and AC power lines. Triggered when rms
value crosses set value in UP or DOWN direction
Period: When rising or falling edge of set voltage does not fall
within cycle range
■
PC Software Specifications
Wave Viewer (Wv) Software (Application disk CD-R, bundled accessory)
• Simple display of waveform file
• Text conversion: convert binary data file to text format, with
selectable space or tab separators in addition to CSV, and speci-
fiable section, thinning available
• Display format settings: scroll functions, enlarge/reduce dis-
Equivalent to 0.25% when full scale is set to 10 divisions
Level setting resolution
Pattern trigger: 1, 0, or × (disregard), logical product (AND) or
logical sum (OR) set for 4 channels
Trigger types
(logic)
Functions
OFF, setting range 0.1 to 10.0 DIV in 9 stages (MEM, REC & MEM
function), ON/OFF (REC function)
Trigger filter
(analog/logic)
play, display channel settings
• Others: voltage value trace function, jump to cursor/trigger
position function
Pre-trigger function to capture pre- and post-trigger waveform,
trigger output (active Low with φ 3.5mm mini-jack and open
collector 5 voltage output), Start & stop trigger in REC function
Other functions
Windows 95/98/Me, Windows NT 4.0 (SP3 or later), 2000, XP
Compatible PC operating systems