7
Output timing of evaluation result
- How it works -
External control connectors insulated from chassis
External DC power can be 12V or 24V, and control signal
input also operates at this voltage. Using the internal isolated
5V power supply for control is also possible. Because the
terminals are isolated from chassis ground, connection of a
sequencer, relay bank, or other device with separate ground
potential presents no problem.
External control connectors provided as standard equipment (rear panel)
Evaluation result output timing
The diagram below shows the evaluation output cycle when performing area evaluation only for a 1kHz sine wave signal. When
auto store is selected, data store processing occurs after the evaluation processing step. The evaluation time and evaluation cycle for
numerical processing depend on the processing type and recording duration, and also on the evaluation waveform.
Trigger mode: continuous
Measurement start
Trigger activated
Trigger activated
Unit operation
Stop Trigger standby Get waveform Evaluation processing Display processing Trigger standby Get waveform
Evaluation time: 70 ms max. (1-channel evaluation), 80 ms max. (2-channel evaluation)
Evaluation cycle: 200 ms max. (1-channel evaluation), 250 ms max. (2-channel evaluation)
Clear previous evaluation result
Evaluation result output
Clear evaluation result
OK_AREA
NG_AREA
Trigger mode: single
Measurement start
Unit operation
Trigger activated
Stop Trigger standby Get waveform Evaluation processing Display processing
Clear previous evaluation result
OK_AREA
NG_AREA
Evaluation result output
Store evaluation result
Measurement stop
Stop
JUDGE_OUT
TRIG_OUT
STORAGE
JUDGE_OUT
BUSY
TRIG_OUT
Setup:
• Time axis range: 100μs/DIV
• Recording length: 10DIV
• Display format: 1screen
• No LAN connection
• A/B cursor display set to OFF
STORAGE
BUSY