欢迎访问ic37.com |
会员登录 免费注册
发布采购

MIDR03-05S05M 参数 Datasheet PDF下载

MIDR03-05S05M图片预览
型号: MIDR03-05S05M
PDF下载: 下载PDF文件 查看货源
内容描述: [Insulation rated for 300VAC Working Voltage]
分类和应用:
文件页数/大小: 4 页 / 453 K
品牌: TOTAL-POWER [ TOTAL POWER INTERNATIONAL ]
 浏览型号MIDR03-05S05M的Datasheet PDF文件第1页浏览型号MIDR03-05S05M的Datasheet PDF文件第2页浏览型号MIDR03-05S05M的Datasheet PDF文件第3页  
MIDR03M SERIES  
DC/DC CONVERTER 3W, Reinforced Insulation, Medical Safety  
Total Power International, Inc.  
Test Setup  
Input Reflected-Ripple Current Test Setup  
Input reflected-ripple current is measured with a inductor Lin (4.7μH) and Cin (220μF, ESR < 1.0at 100 KHz) to simulate source impedance.  
Capacitor Cin, offsets possible battery impedance.  
Current ripple is measured at the input terminals of the module, measurement bandwidth is 0-500 KHz.  
To Oscilloscope  
+Vin  
+Out  
+
+
Lin  
DC / DC  
Converter  
Current  
Probe  
Load  
Battery  
Cin  
-Vin  
-Out  
Peak-to-Peak Output Noise Measurement Test  
Use a Cout 0.33μF ceramic capacitor.  
Scope measurement should be made by using a BNC socket, measurement bandwidth is 0-20 MHz. Position the load between 50 mm and 75 mm from the DC/DC  
Converter.  
Copper Strip  
+Vin  
+Out  
Copper Strip  
+Vin  
+Out  
Com.  
-Out  
Scope  
Cout  
Single Output  
DC / DC  
Converter  
Dual Output  
DC / DC  
Converter  
Resistive  
Load  
Resistive  
Load  
Cout  
Scope  
Copper Strip  
Cout  
Scope  
-Vin  
-Out  
Copper Strip  
-Vin  
Copper Strip  
Technical Notes  
Maximum Capacitive Load  
The MIDR03M series has limitation of maximum connected capacitance at the output. The power module may be operated in current limiting mode during start-up,  
affecting the ramp-up and the startup time. For optimum performance we recommend 220μF maximum capacitive load for dual outputs and 470μF capacitive load  
for single outputs. The maximum capacitance can be found in the data sheet.  
Input Source Impedance  
The power module should be connected to a low ac-impedance input source. Highly inductive source impedances can affect the stability of the power module.  
In applications where power is supplied over long lines and output loading is high, it may be necessary to use a capacitor at the input to ensure startup.  
Capacitor mounted close to the power module helps ensure stability of the unit, it is recommended to use a good quality low Equivalent Series Resistance (ESR <  
1.0at 100 KHz) capacitor of a 4.7μF for the 5V input devices and a 2.2μF for the 12V and 24V devices.  
+
+Vin  
+Out  
+
DC / DC  
Converter  
DC Power  
Source  
Load  
Cin  
-
-Vin  
-Out  
Output Ripple Reduction  
A good quality low ESR capacitor placed as close as practicable across the load will give the best ripple and noise performance. To reduce output ripple, it is  
recommended to use 1.5μF capacitors at the output.  
+
+Vin  
-Vin  
+Out  
+
+Vin  
-Vin  
+Out  
Com.  
-Out  
Single Output  
DC / DC  
Converter  
Load  
Load  
Dual Output  
DC / DC  
Converter  
DC Power  
Source  
Cout  
Cout  
DC Power  
Source  
Cout  
Load  
-
-Out  
-
Thermal Considerations  
Many conditions affect the thermal performance of the power module, such as orientation, airflow over the module and board spacing. To avoid exceeding the  
maximum temperature rating of the components inside the power module, the case temperature must be kept below 90. The derating curves are determined  
from measurements obtained in a test setup.  
Position of air velocity  
probe and thermocouple  
Air Flow  
50mm / 2in  
15mm / 0.6in  
DUT  
Total Power International, Inc  
Toll Free: 877-646-0900  
www.total-power.com