■ Applications of 3193
Support up to Harmonic / Flicker analysis when using the optional 9605 HARMONIC / FLICKER MEASUREMENTS UNIT
.
Graph Display of Harmonics
Voltage, current and power
can be analyzed and displayed
by bargraphs of harmonic
amplitude, content and phase
angle.Voltage, current and power
can be displayed simultaneously
for a single channel, or a single
parameter can be displayed
simultaneously for each of three
channels.
List Display of Harmonics
The harmonic list display
shows the amplitude, pro-
portion, phase angle and
distortion for each harmonic
of voltage, current and power.
Displaying only proportion,
or two parameters simul-
taneously, such as amplitude
and phase angle, is selectable.
Vector Display of Harmonics
The harmonic vector display
shows the voltage, current and
phase angle for each harmonic,
making clear the voltage-
current phase relationship.
Waveform Display
The waveform display shows
one cycle of the voltage and
current waveforms. RMS and
peak values can be displayed
along with voltage and current
waveforms, or voltage and
current waveforms for up to
three channels can be displayed
at the same time.
Flicker Measurement Display
Displays data during measurement
in real-time. Display can also be
switched to D measurement and
"Pst" value.
Monitor Display
The relative "d" voltage change
∆V/V and the instantaneous
flicker value "S(t)" can be
displayed in a time series,
so past variations are clearly
displayed.
9605 Specifications (optional)
:
Every 1 window (except during communications with other devices)
Update rate
: List, graph, vector and waveform
: Harmonic levels; at 45 to 66 Hz
Voltage/current: ±0.5%rdg.±0.05% f.s.
Screen displays
: Installs in the 3193 main unit
Installation
3
Accuracy*
: Single-phase 2- and 3-wire, three-phase 4-wire
: Up to 3 channels within channels 1 to 6, depending
on 3193 wiring mode
Measurement lines
No. of channels
Active power: ±1.0%rdg.±0.1% f.s.
For 45 to 66-Hz fundamental, effective input is 0.1 to 110%
of range, total error of harmonic current when PLL is locked
is 5% of the limit value or within 0.2% of the rated current
of the device under test (no limitation with 9602).
: Floppy disk, RS-232C, GP-IB, printer
Output functions
1
● Harmonic Waveform Analysis Functions (IEC61000-3-2* )
:
Fundamental frequency: 1 to 440 Hz
PLL system (5 to 440 Hz , external clock system (1 to 5 Hz)
Measurement range
2
)
● Flicker Measurement Function (IEC61000-3-3
*
)
: Up to 50th harmonic (with 1 to 250 Hz fundamental)
: 16 cycles (for 40 to 70 Hz fundamental)
: Rectangular tiling (no gap between or overlap of windows)
: 512 points (for 40 to 70 Hz fundamental)
: Up to 4 (current), and up to 3 (voltage)
Orders analyzed
Window width
Windowing type
Amount of data analyzed
Crest factor
:
:
Fundamental frequency; 45 to 66 Hz, PLL synchronization system
dc (relative constant voltage change), d max (max. relative constant
Measurement range
Analysis items
voltage change), d(t)200ms (relative voltage change per time), P0.1/
P1s/ P3s/ P10s and 30s (cumulative probability), Pst (short-term flicker
value), Plt (long-term flicker value)
:
Harmonic level, percentage and phase angle of each order
Measurement items
of harmonic wave for each of voltage, current and power.
Total up to 50th harmonic (of 40 to 70 Hz fundamental) for
voltage, current and power. Total harmonic distortion for
voltage and current (THD-F and THD-R)
:
:
Measured value, CPF, Pst list, Monitor
Screen displays
3
Accuracy*
Within ±5% rdg. of limit on the limit curve line (Pst = 1) specified
for voltage fluctuation in IEC60868-0 (d measurement is the same)
RMS voltage 0.5%rdg. 0.05%f.s. (45Hz~66Hz)
*2. Decision function not included.
Measurement of voltage, current, active power, peak voltage
and peak current values of the fundamental
±
±
*3. The reading accuracy of the input unit must combined with the analysis accuracy shown above. When
used with a clamp sensor, accuracy and frequency characteristics of the clamp must be added to the
analysis accuracy above.
*1. The 9605 does not support limit values, so pass/fail decisions based on limit values, classification
of special waveforms (A to D) and 2.5-minute measurement function for measuring transient harmonic
currents are not supported.