DC current monitor accuracy
±[1% + (0.5 + |V
mon
|/10000)mA] (at 23 ±5 °C)
±[2% + (1.0 + |V
mon
|/5000)mA] (beyond 23 ±5 °C)
25
Ω
(nominal)
Output impedance
NOTE
V
mon
:
I
mon
:
Z
d
=
Z
d
=
Z
d
Z
d
Z
d
dc voltage bias monitor reading value [mV]
dc current bias monitor reading value [mA]
0.3 (at four-terminal pair port of the 4294A, adapter setup: NONE)
2.0 (at 3.5 mm port of the 42941A, adapter setup: 42941A Impedance
Probe)
= 0.5 (at 7-mm port of the 42942A, adapter setup: 42942A Terminal
Adapter)
= 1.0 (at measurement port of the 16048G, adapter setup: four-terminal
pair 1 m)
= 1.5 (at measurement port of the 16048H, adapter setup: four-terminal
pair 2 m)
Figure 1. DC Bias Range (SPC)
Sweep Characteristics
Available sweep parameters
Frequency, Signal voltage, Signal current, DC bias voltage,
DC bias current
Linear, Log
1
, List
1
, Zero-span, Manual
Available for all sweep types
Up sweep, Down sweep
2 to 801 points
Sweep type
Manual sweep
Sweep direction
Number of measurement points
List Sweep
Available setup parameters for
each segment
Sweep frequency range, Number of measurement points, Signal
level (voltage or current), DC bias (voltage or current), Measure-
ment bandwidth, Point averaging factor
1. Frequency sweep only.
5