DC Bias Function
DC voltage bias
Range
Resolution
Accuracy
0 to ±40 V (see Figure 1)
1 mV
±[0.1% + (5 + 30 X |I
mon
|) mV](at 23 ±5 °C)
±[0.2% + (10 + 30 X |I
mon
|) mV](beyond 23 ±5 °C)
DC current bias
Range
Resolution
Accuracy
0 to ±100 mA (see Figure 1)
40 µA
±[2% + (0.2 + |V
mon
|/20) mA] (at 23 ±5 °C)
±[4% + (0.4 + |V
mon
|/20) mA] (beyond 23 ±5 °C)
DC voltage bias at constant
voltage mode
Range
Resolution
Accuracy
0 to ±40 V (see Figure 1)
1 mV
±[0.5% + (5 +
Z
d
X |I
mon
|) mV] (at 23 ±5 °C, typical)
±[1.0% + (10 +
Z
d
X |I
mon
|) mV] (beyond 23 ±5 °C, typical)
DC current bias at constant current mode
Range
Resolution
Accuracy
0 to ±100 mA (see Figure 1)
40 µA
±[1% + (0.5 + |V
mon
|/10000) mA] (at 23 ±5 °C, typical)
±[2% + (1.0 + |V
mon
|/5000) mA] (beyond 23 ±5 °C, typical)
DC bias monitor
DC voltage range
DC voltage accuracy
(Same as the dc voltage bias setting range)
±[0.2% + (5 +
Z
d
X |I
mon
|) mV] (at 23 ±5 °C)
±[0.4% + (10 +
Z
d
X |I
mon
|) mV] (beyond 23 ±5 °C)
(Same as the dc current bias setting range)
DC current range
4