 浏览型号CMA3000-A01PWB的Datasheet PDF文件第6页浏览型号CMA3000-A01PWB的Datasheet PDF文件第7页浏览型号CMA3000-A01PWB的Datasheet PDF文件第8页浏览型号CMA3000-A01PWB的Datasheet PDF文件第9页浏览型号CMA3000-A01PWB的Datasheet PDF文件第11页浏览型号CMA3000-A01PWB的Datasheet PDF文件第12页浏览型号CMA3000-A01PWB的Datasheet PDF文件第13页浏览型号CMA3000-A01PWB的Datasheet PDF文件第14页  
CMA3000-A0X Series  
5
Data sheet references  
5.1 Offset  
CMA3000's offset will be calibrated in X = 0 g, Y = 0 g, and Z = +1 g (Z measuring axis is parallel to  
earth’s gravitation) position, see Figure 8.  
Z-axis in +1 g  
position  
Earth’s  
gravitation  
Y
X
Pin #1  
Figure 8. CMA3000 offset (0 g) position.  
5.1.1 Offset calibration error  
Offset calibration error is the difference between the sensor's actual output reading and the nominal  
output reading in calibration conditions. Error is calculated by  
Equation 1  
OutputX axis Output  
OffsetX axisCalibEr  
=
1000,  
Sens  
where OutputX-axisCalibEr is sensor’s X-axis calibration error in [mg], OutputX-axis is sensor’s X-axis  
output reading [counts], Output is sensor’s nominal output in 0 g position and Sens sensor’s nominal  
sensitivity [counts/g].  
5.1.2 Offset temperature error  
Offset temperature error is the difference between the sensor's output reading in different  
temperatures and the sensor’s calibrated offset value at room temperature. Error is calculated by  
Equation 2  
OutputX axis @T OutputX axis @ RT  
OffsetX axisTempEr @T  
=
1000 ,  
Sens  
where OutputX-axisTempEr@T is sensor’s X-axis temperature error in [mg] in temperature T, OutputX-axis@T  
is sensor’s X-axis output reading [counts] in temperature T, OutputX-axis@T X-axis output reading  
[counts] at room temperature RT and Sens sensor’s nominal sensitivity [counts/g]. Sensor is in 0 g  
position for every measurement point.  
5.2 Sensitivity  
During sensitivity calibration, the sensor is placed in ±1 g positions having one of the sensor’s  
measuring axis at a time parallel to the earth’s gravitation, see Figure 9.  