欢迎访问ic37.com |
会员登录 免费注册
发布采购

71M6515H-IGTW1 参数 Datasheet PDF下载

71M6515H-IGTW1图片预览
型号: 71M6515H-IGTW1
PDF下载: 下载PDF文件 查看货源
内容描述: 截至10ppmC精密超稳定的电压基准数字温度补偿 [Up to 10ppmC precision ultra-stable voltage reference Digital temperature compensation]
分类和应用: 温度补偿
文件页数/大小: 60 页 / 827 K
品牌: TERIDIAN [ TERIDIAN SEMICONDUCTOR CORPORATION ]
 浏览型号71M6515H-IGTW1的Datasheet PDF文件第24页浏览型号71M6515H-IGTW1的Datasheet PDF文件第25页浏览型号71M6515H-IGTW1的Datasheet PDF文件第26页浏览型号71M6515H-IGTW1的Datasheet PDF文件第27页浏览型号71M6515H-IGTW1的Datasheet PDF文件第29页浏览型号71M6515H-IGTW1的Datasheet PDF文件第30页浏览型号71M6515H-IGTW1的Datasheet PDF文件第31页浏览型号71M6515H-IGTW1的Datasheet PDF文件第32页  
71M6515H  
A Maxim Integrated Products Brand  
Energy Meter IC  
DATA SHEET  
JULY 2011  
Subtract 8 from 7:  
9.  
E60 E300 = 2AXV AXI tan(60)sin(φS )  
use equation 5:  
E0 + E180 + 2  
cos(φS )  
10.  
E60 E300  
=
tan(60)sin(φS )  
11.  
E60 E300 = (E0 + E180 + 2) tan(60) tan(φS )  
(E60 E300  
tan(60)(E0 + E180 + 2)  
)
φS = tan1  
12.  
Now that we know the AXV, AXI, and φS errors, we calculate the new calibration voltage gain coefficient from the previous ones:  
CAL _V  
CAL _VNEW  
=
AXV  
We calculate PHADJ from φS, the desired phase lag:  
tan(φS )  
[
1+ (129 )2 2(129 )cos(2πf0T)  
]
PHADJ = 220  
(129 )sin(2πf0T) tan(φS )  
[
1(129 )cos(2πf0T)  
]
Finally, we calculate the new calibration current gain coefficient, including compensation for a slight gain increase in the phase  
calibration circuit.  
CAL _ I  
1
CAL _ INEW  
=
220 PHADJ(2 + 220 PHADJ 2(129 )cos(2πf0T))  
AXI  
1+  
12(129 )cos(2πf0T) + (129 )2  
Alternative Calibration Procedures  
It is possible to implement a fast calibration based on only one measurement with a zero-degree load angle. Details can be  
found in the TERIDIAN Application Note AN_651X_022 (Calibration Procedures).  
Calibration Procedure for Rogowski Sensor  
Rogowski coils generate an output signal that is the derivative of the input current. The 6515H Rogowski module implemented  
in the Rogowski CE image digitally compensates for this effect and has the usual gain and phase calibration adjustments.  
Additionally, calibration adjustments are provided to eliminate voltage coupling from the sensor input.  
Current sensors built from Rogowski coils have relatively high output impedances that are susceptible to capacitive coupling  
from the large voltages present in the meter. The most dominant coupling is usually capacitance between the primary of the  
coil and the coil’s output. This coupling adds a component proportional to the derivative of voltage to the sensor output. This  
effect is compensated by the voltage coupling calibration coefficients.  
As with the CT procedure, the calibration procedure for Rogowski sensors uses the meter’s display to calibrate the voltage  
path and the pulse outputs to perform the remaining energy calibrations. The calibration procedure must be performed to each  
phase separately, making sure that the pulse generator is driven by the accumulated real energy for just that phase. In other  
words, the pulse generator input should be set to WhA, WhB, or WhC, depending on the phase being calibrated. The IC has to  
be configured for Rogowski mode (IMAGE=01). In preparation of the calibration, all calibration parameters are set to their  
default values. VMAX and IMAX are set to reflect the system design parameters. WRATE and PULSE_SLOW, PULSE_FAST are  
adjusted to obtain the desired Kh.  
Page: 28 of 60  
© 20052011 Teridian Semiconductor Corporation  
1.6  
 复制成功!