SPꢀCIFICATIONS
HDO4022
HDO4024
HDO4032
HDO4034
HDO4054
HDO4104
Vertical
Bandwidth (@ 50Ω)
Rise time
Input Channels
Vertical Resolution
Sensitivity
DC Gain Accuracy
BW Limit
200 MHz
1.75 ns typical
350 MHz
1 ns typical
500 MHz
700 ps typical
4
1 GHz
450 ps typical
4
2
4
2
4
12-bits; up to 15-bits with enhanced resolution (ERES)
50 Ω: 1mV/div - 1 V/div; 1 MΩ: 1 mV/div - 10 V/div
(0.5ꢀ) Full Scale, offset at 0 V
20 MHz, 200 MHz
Maximum Input Voltage
Input Coupling
Input Impedance
Offset Range
50 Ω: 5 Vrms; 1 MΩ: 400 V max (DC + Peak AC ≤ 10 kHz)
50 Ω: DC, GND; 1 MΩ: AC, DC, GND
50 Ω 2.0ꢀ, 1 MΩ 2.0ꢀ ꢁꢁ 15 pF
50 Ω: 1 mV - 4.95 mV: 1.6 V, 5 mV - 9.9 mV: 4 V, 10 mV - 19.8 mV: 8 V, 20 mV - 1 V: 10 V
1 MΩ: 1 mV - 4.95 mV: 1.6 V, 5 mV - 9.9 mV: 4 V, 10 mV - 19.8 mV: 8 V, 20 mV - 100 mV: 16 V,
102 mV - 198 mV: 80V, 200 mV - 1 V: 160 V, 1.02 V -10 V: 400 V
Offset Accuracy
(1.0ꢀ of offset value + 0.5ꢀFS + 0.02ꢀ of max offset + 1 mV)
Acquisition
Sample Rate (Single-shot)
Sample Rate (Repetitive)
Record Length
2.5 GS/s
125 GS/s
Standard -STD: 12.5 Mpts/ch (all channels) 25 Mpts (interleaved)
Optional -L:
25 Mpts/ch (all channels), 50 Mpts (interleaved)
Acquisition Modes
Real Time, Roll, RIS (Random Interleaved Sampling),
Sequence (Segmented Memory up to 10,000 segments with 1μs intersegment time)
20 ps/div–10000 s/div; RIS available at ≤ 10 ns/div; Roll Mode available at ≥ 100 ms/div and ≤ 5 MS/s
2.5 ppm for 5 to 40C + 1.0 ppm/year from calibration
Timebase Range
Timebase Accuracy
Trigger System
Modes
Auto, Normal, Single, Stop
Sources
Any input channel, External, Ext/10, or line; slope and level unique to each source (except for line trigger)
Coupling
DC, AC, HFREJ, LFREJ
Pre-trigger Delay
Post-trigger Delay
Hold-off
0-100ꢀ of full scale
0-10,000 Divisions
2ns up to 20s or 1 to 1,000,000,000 events
Internal Trigger Level Range
External Trigger Level Range
Trigger Types
4.1 Divisions
Ext: 400mV, Ext/10: 4V
Edge, Glitch, Width, Logic (Pattern), TV (NTSC, PAL, SECAM, HDTV–720p, 1080i, 1080p), Runt, Slew Rate,
Interval (signal or Pattern), Dropout, Qualified (State or Edge)
Measure, ꢃoom and Math Tools
Measurement Parameters
Up to 8 of the following parameters can be calculated at one time on any waveform: Amplitude, Area, Base (Low),
Delay, Delta Period @ Level, Delta Time @ Level, Duty, Duty @ Level, Edge @ Level, Fall Time (90ꢀ–10ꢀ),
Fall Time (80ꢀ–20ꢀ), Frequency, Frequency @ level, Maximum, Mean, Minimum, Overshoot+, Overshoot-,
Peak-Peak, Period, Period @ Level, Phase, Rise Time (10ꢀ–90ꢀ), Rise Time (20ꢀ–80ꢀ), RMS, Skew,
Standard Deviation, Time @ Level, Top (High), Width+, Width-. Statistics and Histicons can be added to any
measurement and all measurements can be gated.
Zooming
Math Functions
Use front panel QuickZoom button, or use touch screen or mouse to draw a box around the zoom area.
Functions include Sum, Difference, Product, Ratio, Absolute Value, Averaging (summed and continuous), Derivative,
Envelope, Enhanced Resolution (to 15-bits), Floor, Integral, Invert, Reciprocal, Rescale (change scale and units), Roof,
Square, Square Root, Trend, Zoom and FFT (up to 1 Mpts with power spectrum output and rectangular, VonHann, and
FlatTop windows). 2 dual operator math functions may be defined at a time.
14