欢迎访问ic37.com |
会员登录 免费注册
发布采购

S2008V12V 参数 Datasheet PDF下载

S2008V12V图片预览
型号: S2008V12V
PDF下载: 下载PDF文件 查看货源
内容描述: 晶闸管产品目录 [Thyristor Product Catalog]
分类和应用:
文件页数/大小: 224 页 / 2673 K
品牌: TECCOR [ TECCOR ELECTRONICS ]
 浏览型号S2008V12V的Datasheet PDF文件第149页浏览型号S2008V12V的Datasheet PDF文件第150页浏览型号S2008V12V的Datasheet PDF文件第151页浏览型号S2008V12V的Datasheet PDF文件第152页浏览型号S2008V12V的Datasheet PDF文件第154页浏览型号S2008V12V的Datasheet PDF文件第155页浏览型号S2008V12V的Datasheet PDF文件第156页浏览型号S2008V12V的Datasheet PDF文件第157页  
AN1006  
6
Testing Teccor Semiconductor Devices  
Using Curve Tracers  
through several values, and a different trace is drawn on each  
sweep to generate a family of curves.  
Introduction  
One of the most useful and versatile instruments for testing semi-  
conductor devices is the curve tracer (CT). Tektronix is the best  
known manufacturer of curve tracers and produces four basic  
models: 575, 576, 577 and 370. These instruments are specially  
adapted CRT display screens with associated electronics such  
as power supplies, amplifiers, and variable input and output func-  
tions that allow the user to display the operating characteristics of  
a device in an easy-to-read, standard graph form. Operation of  
Tektronix CTs is simple and straightforward and easily taught to  
non-technical personnel. Although widely used by semiconductor  
manufacturers for design and analytical work, the device con-  
sumer will find many uses for the curve tracer, such as incoming  
quality control, failure analysis, and supplier comparison. Curve  
tracers may be easily adapted for go-no go production testing.  
Tektronix also supplies optional accessories for specific applica-  
tions along with other useful hardware.  
Limitations, Accuracy, and Correlation  
Although the curve tracer is a highly versatile device, it is not  
capable of every test that one may wish to perform on semicon-  
ductor devices such as dv/dt, secondary reverse breakdown,  
switching speeds, and others. Also, tests at very high currents  
and/or voltages are difficult to conduct accurately and without  
damaging the devices. A special high-current test fixture avail-  
able from Tektronix can extend operation to 200 A pulsed peak.  
Kelvin contacts available on the 576 and 577 eliminate inaccu-  
racy in voltage measured at high current (VTM) by sensing voltage  
drop due to contact resistance and subtracting from the reading.  
Accuracy of the unit is within the published manufacturer’s speci-  
fication. Allow the curve tracer to warm up and stabilize before  
testing begins. Always expand the horizontal or vertical scale as  
far as possible to increase the resolution. Be judicious in record-  
ing data from the screen, as the trace line width and scale resolu-  
tion factor somewhat limit the accuracy of what may be read.  
Regular calibration checks of the instrument are recommended.  
Some users keep a selection of calibrated devices on hand to  
verify instrument operation when in doubt. Re-calibration or  
adjustment should be performed only by qualified personnel.  
Often discrepancies exist between measurements taken on  
different types of instrument. In particular, most semiconductor  
manufacturers use high-speed, computerized test equipment to  
test devices. They test using very short pulses. If a borderline  
unit is then measured on a curve tracer, it may appear to be out  
of specification. The most common culprit here is heat. When a  
semiconductor device increases in temperature due to current  
flow, certain characteristics may change, notably gate character-  
istics on SCRs, gain on transistors, leakage, and so on. It is very  
difficult to operate the curve tracer in such a way as to eliminate  
the heating effect. Pulsed or single-trace operation helps reduce  
this problem, but care should be taken in comparing curve tracer  
measurements to computer tests. Other factors such as stray  
capacitances, impedance matching, noise, and device oscillation  
also may create differences.  
Tektronix Equipment  
Although Tektronix no longer produces curve tracer model 575,  
many of the units are still operating in the field, and it is still an  
extremely useful instrument. The 576, 577 and 370 are current  
curve tracer models and are more streamlined in their appear-  
ance and operation. The 577 is a less elaborate version of the  
576, yet retains all necessary test functions.  
The following basic functions are common to all curve tracers:  
Power supply supplies positive DC voltage, negative DC volt-  
age, or AC voltage to bias the device. Available power is varied  
by limiting resistors.  
Step generator supplies current or voltage in precise steps to  
control the electrode of the device. The number, polarity, and  
frequency of steps are selectable.  
Horizontal amplifier displays power supply voltage as applied  
to the device. Scale calibration is selectable.  
Vertical amplifier displays current drawn from the supply by  
the device. Scale calibration is selectable.  
Curve tracer controls for beam position, calibration, pulse opera-  
tion, and other functions vary from model to model. The basic  
theory of operation is that for each curve one terminal is driven  
with a constant voltage or current and the other one is swept with  
a half sinewave of voltage. The driving voltage is stepped  
©2002 Teccor Electronics  
Thyristor Product Catalog  
AN1006 - 1  
http://www.teccor.com  
+1 972-580-7777  
 复制成功!