85-264 V
12/24/36/48V Semi-reg
500 W Series
Medical
AC Input:
DC Output:
Power:
RMS
.
Grade:
Technical Specification
MECHANICAL DRAWINGS (1 Module Encased Version — E Package Type)
Top
Right
Bottom
Front
Back
NOTES (applies to all mechanicals)
1)ꢀRecommendedꢀscrewꢀtighteningꢀtorqueꢀofꢀ6ꢀin.lbs
2)ꢀUndimensionedꢀcomponentsꢀareꢀshownꢀforꢀvisualꢀreferenceꢀonly
3)ꢀAllꢀdimensionsꢀinꢀinchesꢀ[mm]
ꢀ ꢀ Tolerances:ꢀ x.xxꢀinꢀ±ꢀ0.02
ꢀ ꢀ
ꢀ
ꢀ
x.xxxꢀinꢀ±ꢀ0.010
Leakage Currents
AC Leakage Current
from Input to Earth
Normal Open Neutral
AC Leakage Current
from Output to Earth
Normal
Condition
Open Earth
Fault
AC Backdrive
AC Line Connection
Model
Condition
Fault
Fault
240 V L-N, 1 phase
125 µA
65 µA
250 µA
130 µA
AQ . . BF . .
AQ . . CF . .
2 µA
2 µA
36 µA
6 µA
125 µA
18 µA
ACuQor Typical at
264 Vac 60 Hz input
ACuQor Typical at
110% nominal input
voltage 60 Hz
208 V L-L, 120 V L-N, 1 of 3 phases
240 V L-N-L, 120 V L-N, split phase
65 µA
130 µA
For convenience, the following tables show limits allowed by various standards:
AC Leakage Current
from Input to Earth
Normal Open Neutral
AC Leakage Current
from Output to Earth
Normal
Condition
Open Earth
Fault
AC Backdrive
Fault
Standard
Contact Type
Condition
Fault
IEC60601-1
NFPA 99 2005
IEC60950
500 µA
300 µA
3500 µA
1000 µA
BF
CF
100 µA
10 µA
500 µA
50 µA
5000 µA
50 µA
Maximum Allowed per
IEC60601-1
Maximum Allowed
per Standard
—
—
Product # AQ0500MUxx Series
Phone 1-888-567-9596
www.synqor.com
Doc.# 005-005046 Rev. 1
09/04/08
Page 5