Technical
Top
Specification
Right
ELI
PR
M
ARY
IN
Bottom
AC Input:
85-264 V
RMS
DC Output:
12/24/36/48V
Semi-reg
.
Power:
500 W Series
Grade:
Medical
Back
MECHANICAL DRAWINGS (1 Module Encased Version — E Package Type)
Top
Right
Bottom
Front
Back
NOTES (applies to all mechanicals)
1) Recommended screw tightening torque of 6 in.lbs
2) Undimensioned components are shown for visual reference only
3) All dimensions in inches [mm]
Tolerances: x.xx in ± 0.02
x.xxx in ± 0.010
Leakage Currents
AC Leakage Current
from Input to Earth
AC Line Connection
Normal Open Neutral
Condition
Fault
125 µA
65 µA
65 µA
250 µA
130 µA
130 µA
AC Leakage Current
from Output to Earth
ACuQor Typical at
264 Vac 60 Hz input
Model
AQ . . BF . .
AQ . . CF . .
Normal
Condition
2 µA
2 µA
Open Earth
Fault
36 µA
6 µA
AC Backdrive
Fault
125 µA
18 µA
240 V L-N, 1 phase
ACuQor Typical at
110% nominal input 208 V L-L, 120 V L-N, 1 of 3 phases
voltage 60 Hz
240 V L-N-L, 120 V L-N, split phase
For convenience, the following tables show limits allowed by various standards:
AC Leakage Current
from Input to Earth
Standard
IEC60601-1
Maximum Allowed
per Standard
NFPA 99 2005
IEC60950
Normal Open Neutral
Condition
Fault
500 µA
300 µA
3500 µA
1000 µA
—
—
AC Leakage Current
from Output to Earth
Maximum Allowed per
IEC60601-1
Contact Type
BF
CF
Normal
Condition
100 µA
10 µA
Open Earth
Fault
500 µA
50 µA
AC Backdrive
Fault
5000 µA
50 µA
Product # AQ0500MUxx Series
Phone 1-888-567-9596
www.synqor.com
Doc.# 005-005046 Rev. 1
09/04/08
Page 5