欢迎访问ic37.com |
会员登录 免费注册
发布采购

AQ0300MU24SCCF 参数 Datasheet PDF下载

AQ0300MU24SCCF图片预览
型号: AQ0300MU24SCCF
PDF下载: 下载PDF文件 查看货源
内容描述: 医疗级AC / DC电源,带PFC [Medical Grade AC/DC Power Supply With PFC]
分类和应用: 医疗功率因数校正
文件页数/大小: 13 页 / 2708 K
品牌: SYNQOR [ SYNQOR WORLDWIDE HEADQUARTERS ]
 浏览型号AQ0300MU24SCCF的Datasheet PDF文件第1页浏览型号AQ0300MU24SCCF的Datasheet PDF文件第2页浏览型号AQ0300MU24SCCF的Datasheet PDF文件第3页浏览型号AQ0300MU24SCCF的Datasheet PDF文件第5页浏览型号AQ0300MU24SCCF的Datasheet PDF文件第6页浏览型号AQ0300MU24SCCF的Datasheet PDF文件第7页浏览型号AQ0300MU24SCCF的Datasheet PDF文件第8页浏览型号AQ0300MU24SCCF的Datasheet PDF文件第9页  
85-264 V  
12/24/36/48V Semi-reg  
300 W Series  
Medical  
AC Input:  
DC Output:  
Power:  
RMS  
.
Grade:  
Technical Specification  
EFFICIENCY, DERATING, AND VOUT DROOP CURVES  
100  
95  
90  
85  
80  
75  
54  
52  
50  
48  
46  
44  
42  
115 Vrms  
70  
230 Vrms  
65  
0
100  
200  
300  
400  
0
1
2
3
4
5
6
7
Output Power (W)  
Output Load (A)  
Figure 8: 48 V  
droop characteristic.  
Figure 7: 48 V  
efficiency curves.  
OUT  
OUT  
500  
400  
300  
200  
100  
0
400  
300  
200  
100  
0
Transient  
Fan Enabled  
Fan Disabled  
Continuous  
132  
170  
264  
85  
0
10  
20  
30  
40  
50  
60  
70  
80  
Input AC Voltage (VRM S  
)
Ambient Temperature (ºC)  
Figure 9: Continuous power derating curve in natural convection.  
Figure 10: Rated output power vs Input AC Voltage.  
Leakage Currents  
AC Leakage Current  
from Input to Earth  
Normal Open Neutral  
AC Leakage Current  
from Output to Earth  
Normal  
Condition  
Open Earth  
Fault  
AC Backdrive  
Fault  
AC Line Connection  
Model  
Condition  
Fault  
240 V L-N, 1 phase  
125 µA  
65 µA  
250 µA  
130 µA  
AQ . . BF . .  
AQ . . CF . .  
2 µA  
2 µA  
36 µA  
6 µA  
125 µA  
18 µA  
ACuQor Typical at  
264 Vac 60 Hz input  
ACuQor Typical at  
110% nominal input  
voltage 60 Hz  
208 V L-L, 120 V L-N, 1 of 3 phases  
240 V L-N-L, 120 V L-N, split phase  
65 µA  
130 µA  
For convenience, the following tables show limits allowed by various standards:  
AC Leakage Current  
from Input to Earth  
Normal Open Neutral  
AC Leakage Current  
from Output to Earth  
Normal  
Condition  
Open Earth  
Fault  
AC Backdrive  
Fault  
Standard  
Contact Type  
Condition  
Fault  
IEC60601-1  
NFPA 99 2005  
IEC60950  
500 µA  
300 µA  
3500 µA  
1000 µA  
BF  
CF  
100 µA  
10 µA  
500 µA  
50 µA  
5000 µA  
50 µA  
Maximum Allowed per  
IEC60601-1  
Maximum Allowed  
per Standard  
Product # AQ0300MUxx Series  
Phone 1-888-567-9596  
www.synqor.com  
Doc.# 005-005044 Rev. A  