GPC200 Commercial/GPM200 Medical 200 Watt Multiple Output
Commercial
Model
Medical
Model
Output
Minimum
Output
Maximum (B)
Output
Maximum (C)
Total
Regulation (A)
Output No.
Output
Output Peak
Noise P-P
GPC200A
GPC200B
GPC200D
GPC200E
GPC200F
GPM200A
GPM200B
GPM200D
GPM200E
GPM200F
1
2
3
4
+5 V
+12 V
-12 V
4 A
0 A
0 A
0 A
18 A
5 A
1 A
3 A
26 A
8 A
1.2 A
4 A
32 A
12 A
1.5 A
4.5 A
50 mV
120 mV
120 mV
120 mV
2%
2%
3%
3%
12 V (D)
1
2
3
4
+ 5 V
+12 V
-5 V
4 A
0 A
0 A
0 A
18 A
5 A
1 A
3 A
26 A
8 A
1.2 A
4 A
32 A
12 A
1.5 A
4.5 A
50 mV
120 mV
50 mV
2%
2%
3%
3%
12 V (D)
120 mV
1
2
3
4
+5 V
+24 V
-12 V
4 A
0 A
0 A
0 A
18 A
3 A
1 A
3 A
26 A
5 A
1.2 A
4 A
32 A
6.5 A
1.5 A
4.5 A
50 mV
240 mV
120 mV
120 mV
2%
2%
3%
3%
12 V (D)
1
2
3
4
+5 V
+24 V
-15 V
4 A
0 A
0 A
0 A
18 A
3 A
1 A
3 A
26 A
5 A
1.2 A
4 A
32 A
6.5 A
1.5 A
4.5 A
50 mV
240 mV
150 mV
150 mV
2%
2%
3%
3%
15 V (D)
1
2
3
4
+5 V
+12 V
-12 V
4 A
0 A
0 A
0 A
18 A
5 A
1 A
3 A
26 A
8 A
1.2 A
4 A
32 A
12 A
1.5 A
4.5 A
50 mV
120 mV
120 mV
50 mV
2%
2%
3%
3%
5 V (D)
A.Total regulation is defined as the maximum deviation from the nominal voltage for all steady-state conditions of initial voltage setting, input line voltage and output load.
B.Ratingsforunrestrictednaturalconvectioncooling;outputs1&2combinedloadnottoexceed22Acontinuous;totalpower-150W.
C.Ratingswith26cfmforcedaircooling;outputs1&2combinedloadnottoexceed32Acontinuous;totalpower-200W.
D.Isolatedoutputwhichmaybereferencedasapositiveornegativevoltage.
GPC200/GPM200 MECHANICAL SPECIFICATIONS
Environmental
Specification
Operating
Non-operating
Temperature (A)
See individual specs
0 to 95% RH
-40 to +85oC
0 to 95% RH
40 gpk
A. Units should be allowed to warm up/operate under non-condensing conditions
before application of power.
B. Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350 to 2000Hz, 3
orthogonal axes.Tested for 10 min./axis operating and 1 hr./axis non-operating.
C. Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction, 3 orthogonal axes,
total 6 shocks.
Humidity (A)
Shock (B)
Altitude
20 gpk
-500 to 10,000 ft
1.5 grms, 0.003 g2/Hz
-500 to 40,000 ft
5 grms, 0.026 g2/Hz
Vibration (C)
Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033
800-235-5929 • 805-486-4565 • FAX 805-487-8911 • www.condorpower.com