欢迎访问ic37.com |
会员登录 免费注册
发布采购

GNT60-15G 参数 Datasheet PDF下载

GNT60-15G图片预览
型号: GNT60-15G
PDF下载: 下载PDF文件 查看货源
内容描述: [AC-DC Regulated Power Supply Module, 65W]
分类和应用: 输出元件
文件页数/大小: 2 页 / 150 K
品牌: SLPOWER [ SL Power Electronics ]
 浏览型号GNT60-15G的Datasheet PDF文件第2页  
GNT60 MEDICAL SERIES
SINGLE OUTPUT POWER SUPPLY
Low Power Medical Power Supplies 65 Watt AC/DC Universal Input Power Supply
Summary:
• 100 - 240 Vac nominal input range
• Overvoltage and short circuit protection
• Approved to UL, CSA and EN Standards
• Class II input available (G2T60)
• RoHS Compliant Model Available (G suffix)
The GNT60 Series is a 65 W universal input AC/DC power
supply in a very small footprint. With the medical approvals
the GNT60 is ideal for a variety of medical device applications
including small single board computers, battery charging,
running small motors, pumps, and solenoids. The series, with
approval to 60601-1 and 60950 standards, improves design-in
time and reduces end system compliance costs.
DC Power Supplies
Specifications:
OUPUT SPECIFICATIONS
Output Power
Total Regulation
Rise time
Transient response
Temperature
Coefficient
Overvoltage
protection
Short Circuit
protection
Cycling/foldback
At turn-on
Main output 50%
0.2 A/μs
Natural Convec-
tion
60 Watts max.(65 W with airflow)
See table
See table
2.0 s max.
3.55% max. dev. 1 μs recovery
to 0.5%
+/-0.03%/oC
108% to 130%
Yes
Ground Leakage
current
Input current
Input fuse
INPUT SPECIFICATIONS
Input Voltage range
Input frequency
Input surge current
Thermistor
limited
264 Vac 60 Hz
120 Vac
230 Vac
F1,F2
Universal input
100-240 nom (85 - 264 max) Vac
55 Hz +/- 10 Hz
37 A max.
300 μA max. (400 μA single fault)
1.2 A
0.7 A
3.15 A
EMC INFORMATION
Conducted emissions
Radiated emissions
Line freq. harmonics
Voltage fluctuations
ESD Air
ESD Contact
Radiated immunity
Fast Transients (EFT)
Line surge immunity
Conducted immunity
Power freq. mag. field
Voltage dip immunity
EN55011
EN55011
EN61000-3-2
EN61000-3-3
EN61000-4-2
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
100 Vac
Level B
Level A
Complies
Complies
8 kV contact
6 kV air
3 V/m
2 kV
1 kV diff /2 kV cm
3 Vrms
3 A/m
5 cycles 40 % Vnom 22% load
25 cycles 70 % Vnom 52% load
0.5 cycles <5% Vnom 88% load
ENVIRONMENTAL SPECIFICATIONS
Thermal Performance
operating ambient
(see ratings chart)
non-operating
0 - 50
o
C
50 -70
o
C ambient
Relative Humidity
Maximum Altitude
Vibration
Shock
non-condensing
operating / non-operating
5 Hz- 500 Hz
per MIL-STD-810E
0-70
o
C
-40 to +85
o
C
Full load. See table.
derate to 50%
5% - 95% RH
10,000 ft. / 40,000 ft. max.
2.5 g rms
516.4 part IV
All specifications are typical at nominal input, full load at 25