Internal Power Supplies
For Medical/Industrial Applications
GB40 Family
PERFORMANCE DATA
Inrush Current
GB40S05K01
GB40S05K01
Fig. 03: INRUSH CURRENT AT 115VAC 5V/5A – CH3: 10A/Div.
Fig. 04: INRUSH CURRENT AT 230VAC 5V/5A – CH3: 10A/Div.
I(inrush) = 16.56A peak.
I²t = 0.27A²s.
I(inrush) = 37.95A peak.
I²t = 1.45A²s.
GB40S12K01
GB40S12K01
Fig. 05: INRUSH CURRENT AT 115VAC 12V/3.4A – CH3: 10A/Div.
Fig. 06: INRUSH CURRENT AT 230VAC 12V/3.4A – CH3: 10A/Div.
I(inrush) = 21.57A peak.
I²t = 0.45A²s.
I(inrush) = 43.72A peak.
I²t = 1.85A²s.
GB40 Application Note v0619
Copyright © 2019 SL Power Electronics Corp. All rights reserved.
Page 6