欢迎访问ic37.com |
会员登录 免费注册
发布采购

G2M65-12 参数 Datasheet PDF下载

G2M65-12图片预览
型号: G2M65-12
PDF下载: 下载PDF文件 查看货源
内容描述: G2M65 - X,其中X是任何数目从12到48 ,它表示输出电压额定值。 [G2M65-X, where X is any number from 12 through 48, which represents the output voltage rating.]
分类和应用:
文件页数/大小: 2 页 / 113 K
品牌: SLPOWER [ SL Power Electronics ]
 浏览型号G2M65-12的Datasheet PDF文件第2页  
SL Power Electronics, Corp.
6050 King Drive., Bldg A
Ventura, CA 93003
Internet: www.slpower..com
INSTALLATION INSTRUCTIONS
G2M65 SERIES
MODEL NUMBERS:
G2M65-X, where X is any number from 12 through 48, which represents the
output voltage rating. May be followed by suffix –XXX where XXX may be any number from 001 thru
999 to indicate value added configurations that have no impact on safety and/or suffix G to indicate
compliance to RoHS.
RATINGS:
Input:
Output:
100-240 V ac, 1.5 A, 50/60 Hz
5.4 A or 65 W maximum or see table for standard output voltage models.
Model
Output
Watts
G2M65-12
G2M65-15
G2M65-24
G2M65-28
G2M65-48
+12 V dc 5.4 A
+15 V dc 4.3 A
+24 V dc 2.7 A
+28 V dc 2.3 A
+48 V dc 1.35 A
65
65
65
65
65
Notes:
1. Maximum ambient temperature for rated output current is 50 °C.
2. Maximum Operating Relative Humidity 96 %, no condensation.
3. Storage: -40 to +85 °C. Units should be allowed to warm-up under non-condensing conditions before application of
power.
CERTIFICATION:
All models are Certified to be in compliance with the applicable requirements of
UL 60601-1, CSA 22.2 No. 601.1, EN 60601-1, and IEC 60601-1.
CLASSIFICATION:
(5.1) Protection against electric shock = Class II
(In accordance with sub-
(5.2) Degree of protection against electric shock = Signal output or intermediate
clause 5 of IEC 60601-1)
(5.3) Protection against harmful ingress of water = Ordinary (no protection)
(5.5) Have not been evaluated for use in the presence of a flammable anaesthetic
mixture with air, oxygen, or nitrous oxide. This evaluation is to be made
on the end equipment by the OEM.
(5.6) Mode of operation = Continuous
SAFETY DECLARATION:
Condor DC Power Supplies, Inc. declares under our sole
responsibility that all models listed above are in conformity with the applicable
requirements of EN 60950-1 following the provisions of the Low Voltage Directive
73/23/EEC.
OUTPUT:
The output must not be connected to Earth ground or to metallic enclosure in the end
application. The output is not acceptable for direct patient connection without additional isolation. The
DC output is SELV under normal and single fault conditions.
OVERVOLTAGE PROTECTION:
The output is monitored for an overvoltage condition. In some
applications where an overvoltage condition could result in a hazard as defined in applicable safety
standards, redundant or additional overvoltage protection may be required. Consult factory for details.
41-35405-0001 Rev. C 6/8/06
Page 1 of 2