 浏览型号NICRIO-9052PXI的Datasheet PDF文件第1页浏览型号NICRIO-9052PXI的Datasheet PDF文件第3页  
High-Speed PCI Interface for CompactRIO
Safety Standards
The cRIO-9052 is designed to meet the requirements of the following
safety standards for electrical equipment for measurement, control,and
laboratory use:
EN 61010-1, IEC 61010-1
UL 61010-1
CAN/CSA-C22.2 No. 61010-1
Note:
For UL and other safety certifications, refer to the product label, or
visit
ni.com/certification.
Specifications
StarFabric Communication
Interface .............................................. 4 LVDS transmit pairs,
4 LVDS receive pairs
Communication rates
Theoretical maximum ..................... 133 MB/s
Typical
1
........................................... 50 MB/s
1
This rate varies depending on your host system specifications, such as processor, RAM, and so on.
Cabling requirements
Cables ............................................. 2 shielded CAT5 Ethernet cables
Maximum difference between
the lengths of the two cables.... 2.54 cm (1 in.)
Maximum cabling length................ 14 m (45.93 ft)
Environmental
The cRIO-9052 is intended for indoor use only. For outdoor use, mount
the CompactRIO system in a suitably rated enclosure.
Operating temperature
(IEC 60068-2-1, IEC 60068-2-2) ...... -40 to 70 °C
Note:
To meet this operating temperature range, follow the guidelines in
the installation instructions for the CompactRIO system.
Storage temperature
(IEC 60068-2-1, IEC 60068-2-2) ......
Ingress protection ...............................
Operating humidity (IEC 60068-2-56).....
Storage humidity (IEC 60068-2-56)........
Maximum altitude...............................
Pollution degree (IEC 60664) ..............
-40 to 85 °C
IP 40
10 to 90% RH, noncondensing
5 to 95% RH, noncondensing
2,000 m
2
Power Requirements
You must use a National Electric Code (NEC) Class 2 power source with
the cRIO-9052.
Recommended power supply.............. 48 W secondary,
18 VDC to 24 VDC
Power consumption
cRIO-9052 only................................ 1.5 W, max
cRIO-9052 supplying power to
8 CompactRIO modules .................. 17 W, max
1
Power supply
On power up ................................... 9 to 35 V
After power up ............................... 6 to 35 V
1
Shock and Vibration
To meet these specifications, you must panel mount the
CompactRIOsystem and affix ferrules to the end of the terminal wires.
Operating vibration
Random (IEC 60068-2-64) ............... 5 g
rms
, 10 to 500 Hz
Sinusoidal (IEC 60068-2-6) ............. 5 g, 10 to 500 Hz
Operating shock (IEC 60068-2-27) ...... 30 g, 11 ms half sine, 50 g,
3 ms half sine,
18 shocks at 6 orientations
This is the maximum amount of power that the entire system should consume
at the power connector.
Note:
The cRIO-9052 powers up only at 9 V or higher, but after it is
powered up, it can run on as little as 6 V.
Physical Characteristics
If you need to clean the device, wipe it with a dry towel.
Screw-terminal wiring. ....................... 12 to 24 AWG copper conductor
wire with 10 mm (0.39 in.) of
insulation stripped from the end
Torque for screw terminals................. 0.5 to 0.6 N
m
(4.4 to 5.3 lb
in.)
Weight.................................................... Approx. 546 g (19.2 oz)
Electromagnetic Compatibility
Emissions ............................................ EN 55011 Class A at 10 m FCC
Part 15A above 1 GHz
Immunity.............................................. Industrial levels per EN
61326:1997 + A2:2001, Table A.1
EMC/EMI............................................. CE, C-Tick, and FCC Part 15
(Class A) Compliant
Note:
For EMC compliance, operate this device with shielded cabling.
Safety
Safety Voltages
Connect only voltages that are within these limits.
V-to-C............................................................ 35 V max, Measurement Category I
Measurement Category I is for measurements performed on circuits not
directly connected to the electrical distribution system referred to as
MAINS voltage. MAINS is a hazardous live electrical supply system that
powers equipment. This category is for measurements of voltages from
specially protected secondary circuits. Such voltage measurements
include signal levels, special equipment, limited-energy parts of equipment,
circuits powered by regulated low-voltage sources, and electronics.
CE Compliance
This product meets the essential requirements of applicable European
directives, as amended for CE marking, as follows:
Low-Voltage Directive (safety) ........... 73/23/EEC
Electromagnetic Compatibility
Directive (EMC)............................... 89/336/EEC
Note:
Refer to the Declaration of Conformity (DoC) for this product for
any additional regulatory compliance information. To obtain the DoC for
this product, visit
ni.com/certification.
BUY ONLINE
at
ni.com
or
CALL (800) 813 3693 (U.S.)