250 MS/s, 125 MHz, 8-Bit Digitizers
External Trigger Channel (TRIG)
Impedance........................................... 1 MΩ || 22 pF
Vertical range...................................... ±5 V
Coupling .............................................. AC, DC
Safety and Compliance
Safety
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 61010-1, CSA 61010-1
Note:
For UL and other safety certifications, refer to the product label or
visit
ni.com/certification,
search by model number or product line, and
click the appropriate link in the Certification column.
Intermodule SMC Synchronization Using
NI-TClk (typical)
Skew.................................................... 500 ps
≤20
ps after manual adjustment
Power Requirements (typical)
PXI
PCI
+3.3 VDC
840 mA
1.6 A
+5 VDC
1.1 A
1.7 A
+12 VDC
250 mA
45 mA
-12 VDC
170 mA
0A
Total Power
13.32 W
14.32 W
Electromagnetic Compatibility
This product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement, control,
and laboratory use:
• EN 61326 EMC requirements; Minimum Immunity
• EN 55011 Emissions; Group 1, Class A
• CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A
Note:
For EMC compliance, operate this device according to
product documentation.
Environment
Operating temperature
2
...................... 0 to 55 °C for PXI-5114,
0 to 45 °C for PCI-5114
(meets IEC-60068-2-1 and
IEC-60068-2-2)
Storage temperature........................... -40 to 71 °C (meets
IEC-60068-2-1 and IEC-60068-2-2)
Relative humidity ................................ 10 to 90%, noncondensing
(meets IEC-60068-2-56)
2
0
CE Compliance
This product meets the essential requirements of applicable European
Directives, as amended for CE marking, as follows:
• 2006/95/EC; Low-Voltage Directive (safety)
• 2004/108/EC; Electromagnetic Compatibility Directive (EMC)
Note:
Refer to the Declaration of Conformity (DoC) for this product for
any additional regulatory compliance information. To obtain the DoC for
this product, visit
ni.com/certification,
search by model number or
product line, and click the appropriate link in the Certification column.
to 45 °C in PXI-101x or PXI-1000/B chassis.
Calibration
Self-calibration.................................... Gain, offset, compensated
1 MΩ attenuator, triggering,
and timing for all input ranges
External calibration interval................ 2 years
Waste Electrical and Electronic Equipment (WEEE)
EU Customers:
At the end of their life cycle, all products must be
sent to a WEEE recycling center. For more information about WEEE
recycling centers and National Instruments WEEE initiatives,
visit
ni.com/environment/weee.htm.
BUY ONLINE
at
ni.com
or
CALL 800 813 3693 (U.S.)
3