欢迎访问ic37.com |
会员登录 免费注册
发布采购

778539-0M 参数 Datasheet PDF下载

778539-0M图片预览
型号: 778539-0M
PDF下载: 下载PDF文件 查看货源
内容描述: 100和50 MHz的可编程VoltageDigital波形发生器/分析仪 [100 and 50 MHz, Programmable-VoltageDigital Waveform Generator/Analyzers]
分类和应用:
文件页数/大小: 6 页 / 548 K
品牌: NI [ NATIONAL INSTRUMENTS CORPORATION ]
 浏览型号778539-0M的Datasheet PDF文件第1页浏览型号778539-0M的Datasheet PDF文件第2页浏览型号778539-0M的Datasheet PDF文件第3页浏览型号778539-0M的Datasheet PDF文件第4页浏览型号778539-0M的Datasheet PDF文件第6页  
100 and 50 MHz, Programmable-Voltage
Digital Waveform Generator/Analyzers
Events (outputs from the NI 655x)
Event types............................................................................. Marker <0..3>, Data Active event, Ready for Start event
Destinations........................................................................... 1. PFI 0 (SMB jack connectors)
2. PFI <1..3> (DDC connector)
3. PXI_TRIG <0..7> (PXI backplane)
Each of the events can be routed to any of the destinations with the exception of the
Data Active event. The Data Active event can only be routed to the PFI channels.
Miscellaneous
Interval for external calibration............................................. 2 years
Onboard Clock Characteristics (only valid when PLL reference source is set to “None”)
Frequency accuracy................................................................ ±100 ppm (typical)
Temperature stability............................................................. ±30 ppm (typical)
Aging...................................................................................... ±5 ppm first year (typical)
Power Requirements
Typical .................................................................................... 21.6 W
Maximum ............................................................................... PXI: 26.5 W; PCI: 27 W
Physical
Dimensions ............................................................................ PXI: Single 3U CompactPCI Slot. PXI Compatible
PCI: 12.6 by 35.5 cm (4.95 by 13.9 in.)
Environment
I/O Panel Connectors
Label
CLK IN
PFI 0
CLK OUT
Digital Data and
Control (DDC)
External Function(s)
Sample Clock, External PLL
Reference Input
Events, Triggers
Exported Sample Clock,
Exported Reference Clock
Digital Data Channels, Exported
Sample Clock, STROBE, Events, Triggers
Connector Type
SMB jack
SMB jack
SMB jack
68-pin VHDCI
Operating temperature .......................................................... PXI: 0 to 55 °C in all NI PXI chassis except the following. 0 to 45 °C when installed
in an NI PXI-1000/B and PXI-101x chassis (meets IEC-60068-2-1 and IEC-60068-2-2).
PCI: 0 to 45 °C
Storage temperature.............................................................. -20 to 70 °C
Relative humidity ................................................................... 10 to 90% relative humidity, noncondensing (meets IEC-60068-2-56)
Storage relative humidity ...................................................... 5 to 95% relative humidity, noncondensing (meets IEC-60068-2-56)
Certifications and Compliance
Safety
NI 655x devices meet the requirements of the following standards for safety and electrical equipment for measurement, control,
and laboratory use: IEC 61010-1, EN 61010-1, UL 61010B-1, CAN/CSA C22.2 No. 61010-1
EMC ....................................................................................... Emissions: EN 55011 Class A at 10 m. FCC Part 15A above 1 GHz Immunity:
EN 61326:1997 + A2:2001, Table 1, CE, C-Tick, and FCC Part 15 (Class A) Compliant
BUY ONLINE
at
ni.com
or
CALL (866) 265 9891 (U.S.)
5