29 | Keysight | U4164A Logic Analyzer Module - Data Sheet  
Environmental and Physical  
Operating environment  
Temperature (nom)  
Humidity (nom)  
0 to +40 °C  
0 to 80% relative humidity at 40 °C  
Altitude  
0 to 3000 m  
Vibration  
Random vibration 5 to 500 Hz, 10 minutes per axis, approximately 0.2 g rms  
Non-operating environment  
Temperature (nom)  
Humidity (nom)  
–40 to +75 °C  
0 to 90% relative humidity at 65 °C  
0 to 15,300 m  
Altitude  
Vibration (in shipping carton)  
Random vibration 5 to 500 Hz  
10 minutes per axis  
Approximately 2.41  
Swept sine resonant search  
5 to 500 Hz  
0.50 g (0-peak)  
5 minute resonant dwell at 4 resonances per axis  
2.34 kg  
Weight  
Definitions for specifications  
Specification (spec): Represents warranted performance of a calibrated instrument that has been stored for a minimum  
of 2 hours within the operating temperature range of 0 to 40 °C, unless otherwise stated, and after a 45 minute warm-up period. The  
specifications include measurement uncertainty. Data represented in this document are specifications unless otherwise noted.  
Typical (typ): Represents characteristic performance, which 80% of the instruments manufactured will meet. This data is not  
warranted, does not include measurement uncertainty, and is valid only at room temperature (approximately 25 °C).  
Nominal (nom): The expected mean or average performance, or an attribute whose performance is by design, such as the 50 Ω  
connector. This data is not warranted and is measured at room temperature (approximately 25 °C).  
Measured (meas): An attribute measured during the design phase for purposes of communicating expected performance, such as  
amplitude drift vs. time. This data is not warranted and is measured at room temperature (approximately 25 °C).  
 复制成功!