29 | Keysight | U4164A Logic Analyzer Module - Data Sheet
Environmental and Physical
Operating environment
Temperature (nom)
Humidity (nom)
0 to +40 °C
0 to 80% relative humidity at 40 °C
Altitude
0 to 3000 m
Vibration
Random vibration 5 to 500 Hz, 10 minutes per axis, approximately 0.2 g rms
Non-operating environment
Temperature (nom)
Humidity (nom)
–40 to +75 °C
0 to 90% relative humidity at 65 °C
0 to 15,300 m
Altitude
Vibration (in shipping carton)
Random vibration 5 to 500 Hz
10 minutes per axis
Approximately 2.41
Swept sine resonant search
5 to 500 Hz
0.50 g (0-peak)
5 minute resonant dwell at 4 resonances per axis
2.34 kg
Weight
Definitions for specifications
Specification (spec): Represents warranted performance of a calibrated instrument that has been stored for a minimum
of 2 hours within the operating temperature range of 0 to 40 °C, unless otherwise stated, and after a 45 minute warm-up period. The
specifications include measurement uncertainty. Data represented in this document are specifications unless otherwise noted.
Typical (typ): Represents characteristic performance, which 80% of the instruments manufactured will meet. This data is not
warranted, does not include measurement uncertainty, and is valid only at room temperature (approximately 25 °C).
Nominal (nom): The expected mean or average performance, or an attribute whose performance is by design, such as the 50 Ω
connector. This data is not warranted and is measured at room temperature (approximately 25 °C).
Measured (meas): An attribute measured during the design phase for purposes of communicating expected performance, such as
amplitude drift vs. time. This data is not warranted and is measured at room temperature (approximately 25 °C).